Pressure sensor

Advertisement

Use this button to re-calibrate the pressure sensor and follow the on-screen instructions to step through the
required process. To perform a single-point calibration, select 'NO' when prompted to proceed with the next
calibration point. Select 'YES' otherwise to complete the two-point calibration.
The user may need to repeat the calibration procedure to fine tune the offset and gain factors when large
changes are made.
The bottom section of the screen displays current readings for both pressure and temperature; this is only
visible when RS232 is selected as sensor output.
For loading a configuration file stored on PC to the
sensor. This button is enabled as soon as a valid
serial number is received. The interface will first
check for a serial number match before proceeding
to data transfer and resetting the sensor.
Use this button to save the current sensor settings
to a PC file. This button is enabled only after
connection to a sensor has been established.
Fully Documented Help Function is available for PS7000 Utility by
clicking the Help button or hitting the F1 key