Pressure sensor

Advertisement

The sensor is assembled and calibrated to the required range using Ruska Digital Pressure controllers which
are externally calibrated in NATA certified laboratories.
The sensor is calibrated at multiple points over its pressure and temperature range
(typically 36 points).
o
The calibration is validated at multiple different points (typically 25 points).
o
Accuracy and linearity is calculated from the validation data.
An extensive range of final calibration and inspection tests are carried out on every sensor.
The sensor is visually inspected and packed, ready for despatch.
The complete calibration records, sensor history and batch number are placed on file and archived.