Areas of application
Upper basket/injector unit O 177/1
Left hand side of basket: holder and 28
jets in silicone holders for the direct
injection of hollow instruments. This
ensures adequate cleaning and
disinfection both internally and
externally.
Right hand side of basket: free e.g. for
E 337 insert for upright instruments.
,
To minimise the risk of injury
from instruments that are sharp at
each end, and when arranging
instruments that have a probe
pointing upwards, instruments
should be loaded starting at the
back and working forwards, and
unloaded starting at the front and
working backwards.
22
Upper basket/injector unit O 177/1
With E 413 insert for 6 thermally stable
suction tubes (e.g. Sirona System).
,
Hollow instruments and tubes
must be dried internally with filtered
compressed air.
This also ensures there are no
blockages.