Service Procedures

cabinet cleaning

Tools/Equipment
Method

Hose replacement

Method
Figure 1. Attaching the
replacement Hose
Press this
tab to
release
hose
Bair Hugger Model 505 Service Manual
Soft cloth, lightly dampened with water and mild detergent.
1.
Disconnect the warming unit from the power source.
2.
Wipe the exterior of the warming unit clean.
CAUTION: Do not use a dripping wet cloth when cleaning the
warming unit. Moisture may seep into the electrical contacts
and damage the components. Also, do not use alcohol or other
solvents to clean the cabinet. Solvents may damage the labels and
other plastic parts.
1.
Press the tab on the blower outlet to release the existing
hose.
2.
Attach the replacement hose by inserting the flange end at
a 45° angle in the grooved blower outlet, then snap the hose
into place. (See Figure 1.)
Insert flange into
groove and snap into
place