normal operating Temperature Testing

Service Frequency
Tools/Equipment
Method
Figure 8. room Temperature vs. High
Temperature output
Warming unit temperature setting should be on HIGH when
conducting this test
45
44
43
42
15 16 17 18 19 20 21 22 23 24 25 26 27
Ambient Room Temperature (°C)
Bair Hugger Model 505 Service Manual
WARNING: Perform all testing and adjustments of warming unit
temperature with a calibrated Model 221 Temperature Test Kit
or a Model 22110 Temperature Test Unit. Arizant Healthcare Inc.
assumes no responsibility for the reliability, safety, or performance
of the Bair Hugger system if warming unit temperature testing or
adjustments are made in any other manner than that described
here. Improper measurement or adjustment of the warming unit's
normal operating temperature could result in patient exposure
to temperatures outside of the indicated range, and may lead to
patient injury.
6-month intervals, or 500 hours of use
Model 221 Temperature Test Kit or Model 22110 Temperature
Test Unit
1.
Insert the open hose end from the warming unit in the port
on the temperature test device (see Figure 6).
NOTE: Do not block the rear air vents on the test device.
2.
Turn the warming unit power ON and select the
HIGH temperature setting. Always begin testing at the
HIGH setting.
3.
Note the ambient room temperature.
4.
After approximately 5 minutes of operation, note the
temperature read-out. Use Figure 8 to correlate ambient
temperature to the warming unit's operating temperature.
1