Air Filter replacement

Tools/Equipment
Service Frequency
Method
Bair Hugger Model 505 Service Manual
The air filter is mounted in the bottom of the warming unit
(See Figure 4).
Phillips-head screwdriver
Every 6 months or 500 hours of use.
1.
Disconnect the warming unit from the power source.
2.
Remove the M6 x 25 pan-head screw on the bottom on the
unit. The filter should slide out easily.
3.
Record the hour meter reading and date on the new air
filter label. The hour meter is located in the lower left
corner of the top warning label (see Figure 4).
4.
Slide the new filter into the base of the unit.
5.
Replace the M6 x 25 pan-head screw on the bottom on the
unit.
6.
Reconnect the warming unit to the power source.
7.
Record the hour meter reading and maintenance action
taken in the Maintenance Log section.
Figure 4. location
of Air Filter and
Hour Meter
Hour meter
Air filter