RUN TIMES

Multiple factors will affect Battery Pack actual run times including, but
not limited to,
device type, whether a power inverter or power converter
is used, pressure setting, and altitude. The use of heated humidification
or heated tubing will significantly reduce expected run times.
Average run times are based on Respironics System One and ResMed S-9
Series PAP devices with no heated humidification or heated tubing.
SINGLE BATTERY CONFIGURATION
Respironics & other 12V devices: 1-3 nights
*ResMed & other 24V devices: 1-2 nights
DUAL BATTERY CONFIGURATION
Respironics & other 12V devices: 3-5 nights
*ResMed & other 24V devices: 2-3 nights
*Use of a power converter or power inverter required.

RESMED DEVICE WARNING

If you are using any device manufactured by ResMed
you must use either
a power converter or power inverter to complete the connection between
the device and the Battery Pack.
A power converter or power inverter must be used because ResMed
devices operate from a 24V direct current and the plugs may run opposite
polarity from other manufacturer devices.
Do not connect the Battery Pack directly to any ResMed model device
as doing so will damage the device.
See pages 16-17 for additional instructions regarding ResMed devices.