Specifications; Initial Preparation - Hanna Instruments HI 98710 Instruction Manual

Printing and logging thermometers
Hide thumbs Also See for HI 98710:
Table of Contents

Advertisement

SPECIFICATIONS

Range
-50.0 to 150.0°C ; -55.0 to 300.0°F
Resolution
0.1°C (-30 to 130°C); 0.2°C outside
0.1°F (-18 to 225°F);
0.2°F (225 to 260°F); 0.3°F outside
Accuracy
±0.4°C (-20 to 120°C); ±0.7°C outside
(@20°C/68°F)
±0.8°F (-4 to 248 °F); ±1.3°F outside
for one year (excluding probe error)
Typical EMC Dev.
±0.4°C; ±0.8°F
Channels
HI 98710 / HI 98810: 1 channel
HI 98740 / HI 98840: 4 channels
Probe
HI 762 series
Printer
Low power impact, 14 characters per line, using
38 mm plain paper (HI 710034)
Printing/Logging
Selectable at
Interval
1, 2, 5, 10, 15, 30, 60, 120 or 180 minutes
Auto-off
Selectable after 5, 10, 15, 30, 45 or 60 minutes
Power supply
4 x 1.5V AA alkaline batteries / 350 hours of use
(without printing and backlight); or 12 Vdc input
Environment
0 to 50°C (32 to 122°F); RH max 95%
Dimensions
220 x 82 x 66 mm (8.7 x 3.2 x 2.6")
Weight
550 g (1.2 lb.)
6

INITIAL PREPARATION

Each meter is supplied complete with batteries. Remove the back
cover, unwrap the batteries and install them while paying attention
to the polarity.
Alternatively, connect the a 12 Vdc voltage adapter to the power
adapter plug.
To prepare the instrument for use, choose the most appropriate
temperature probe(s) for your application (see "Accessories") and
connect it (them) to the the instrument.
All the probes have been precalibrated at the factory and no
calibration is needed.
With the meter facing you, channel #1 is the first connector on the
top left hand side.
To switch the meter on, press the ON/OFF key.
The batteries charge status or "LINE" message
(if external power adapter is connected) will be
displayed on the LCD for a few seconds.
The meter is now ready to operate.
To maximize battery life, the meter is automatically switched off after a
user selectable period of non-use (this feature is enabled and set to 5
minutes by default; it can be disabled or changed through setup
code #40).
If in logging mode, after the period of non-use, the meter will
continue to monitor the temperature periodically at the end of every
logging interval. Only the "LOG" indication will be visible on LCD.
While storing data in memory, during the sleep mode, the reading
will appear briefly on the LCD.
To reactivate the display press the ON/OFF key.
Note: When the use of an alternate function (FNC, CFM and LOT
INFO) is requested, press and hold the ALT key first and then
the second key.