Hanna Instruments HI 98710 Instruction Manual

Hanna Instruments HI 98710 Instruction Manual

Printing and logging thermometers
Hide thumbs Also See for HI 98710:

Advertisement

Instruction Manual
HI 98710 • HI 98810
HI 98740 • HI 98840
Printing and Logging
Thermometers
w w w . h a n n a i n s t . c o m

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HI 98710 and is the answer not in the manual?

Questions and answers

Summary of Contents for Hanna Instruments HI 98710

  • Page 1 Instruction Manual HI 98710 • HI 98810 HI 98740 • HI 98840 Printing and Logging Thermometers w w w . h a n n a i n s t . c o m...
  • Page 2: Table Of Contents

    PRELIMINARY EXAMINATION Dear Customer, Thank you for choosing a HANNA instruments ® product. Please read this instruction manual carefully before using the instru- Remove the instrument from the packing material and examine it ment. carefully to make sure that no damage has occurred during shipping.
  • Page 3: Functional Description

    HI 9200 infrared interface, connected to the computer serial port. The connection is handled with the HI 92000 Windows ® com- patible software from HANNA instruments ® HI 98810 accepts one temperature probe while HI 98840 can be connected to 4 separate probes at the same time.
  • Page 4: Specifications

    SPECIFICATIONS INITIAL PREPARATION Each meter is supplied complete with batteries. Remove the back Range -50.0 to 150.0°C ; -55.0 to 300.0°F cover, unwrap the batteries and install them while paying attention Resolution 0.1°C (-30 to 130°C); 0.2°C outside to the polarity. 0.1°F (-18 to 225°F);...
  • Page 5: Setup Mode

    PASSWORD PROTECTION SETUP MODE Setting of the GLP parameters (calibration alarm time-out, instru- ment ID code, time and date) can be password protected. If password Setup can be used to view data regarding instrument status (e.g. is set to a value different from 0000 (factory setting), the user will be battery charge) or GLP data (e.g.
  • Page 6 cod. 41: Battery test Code Valid values Default cod. 50: Baud rate 06 Timed data delete cod. 60: Release code 10 Show GLP data cod. 70: Celsius or Fahrenheit 11 Calibration alarm time-out On(enabled); Off(disabled) On cod. 99: Pass Code 20 Instrument ID code 0000÷9999 0000...
  • Page 7 Viewing logged data TO RESET PRINTING SAMPLE NUMBER • Press CH/TIME (HI98840 only) or TEMP/TIME (HI98810 only) to (HI98710 and HI98740 only) view data of the selected sample. Data will be displayed in the This feature resets the print on demand sample number to 001. following order: •...
  • Page 8: Taking Temperature Measurements

    HI 98810 - HI 98840 TAKING TEMPERATURE MEASUREMENTS PRINTING / LOGGING FUNCTIONS To prepare the instrument for use, choose the most appropriate tem- Two different modes to print / log data are available: perature probe(s) for your application (see accessories) and connect it 1.
  • Page 9 TO STOP LOGGING mand data area). If the print function is enabled, a printout is also produced providing the following information: In order to stop the recording mode, press a - Date ALT and LOG keys (press ON/OFF first, if b - Sample number meter is in sleep mode).
  • Page 10: Hi98710 / Hi98740 Printing Functions

    TO RETRIEVE LOGGED DATA TO STOP PRINTING Logged data can be viewed on LCD or printed. To view or print logged In order to stop the recording mode, press data see "TO SCAN LOGGED DATA" in the "SETUP MODE" paragraph. ALT and LOG keys (press ON/OFF first, if meter is in sleep mode).
  • Page 11: Good Laboratory Practice (Glp)

    This feature can be enabled and disabled pressing the ALT and HI 92000 Windows ® compatible software by HANNA instruments ® LIGHT keys. The LCD backlight can be disabled in order to save power and it is automatically disabled when battery charge weakens.
  • Page 12: Memory Organization

    PRINTER MAINTENANCE MEMORY ORGANIZATION HI 98810 - HI 98840 only TO CHANGE THE INK CARTRIDGE When printouts become faint, it might be necessary to change the ink Logged data are stored in the internal EEPROM and are retained even cartridge. Contact your Hanna authorized center. if batteries and external power are disconnected.
  • Page 13: Battery Replacement

    BATTERY REPLACEMENT TEMPERATURE PROBES When the batteries are inserted and no power adapter is connected, HI 762A Air probe, 1 m (3.3') cable and white handle the meter can recognize different batteries charge levels. HI 762A/10 Air probe, 10 m (33') cable and white handle 1.
  • Page 14: Un-Housed Ntc Sensors

    HI 76S2-5 NTC thermistor sensor, aluminum tab with mounting UN-HOUSED NTC SENSORS hole 300mm 12" 0.31" DIA 3mm HANNA instruments ® offers a wide line of NTC thermistor sensors that 1.6mm 0.12" 0.06" EPOXY can be custom mounted in almost any situation. Select the sensor 12.7mm...
  • Page 15: Warranty

    WARRANTY All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and main- tained according to instructions. The electrodes and the probes are warranted for a period of six months. This warranty is limited to repair or replacement free of charge.