3
Installation
3.4

Connections

3.4
Connections
Communication
Connect EPS 3501 XL according to the electrical drawings in Appendix A.
Connect the leads from the electrophoresis unit to the output sockets on the EPS 3501
XL front panel (red to red, and black or blue to blue), see Section 1.3. The red lead is the
positive and black or blue is the negative.
Electrical power
Connect the power cord to a grounded power outlet specified in Section 3.1.
Select the appropriate voltage range, 100–120 or 220–240 V, using the switch on the
rear panel.
Local regulation for Great Britain
This appliance must be earthed.
The wires in the mains lead are colored in accordance with the following code:
Color
Green and yellow
Blue
Brown
If the plug provided is unsuitable for your socket outlets, the plug must be cut off and
suitable plug fitted. The cut-off plug should be disposed of and must not be inserted into
any 13 amp socket as this can result in electric shock. The plug or adapter of the
distribution panel should be provided with 13 amps fuse. As the colors of the wires in
the mains lead of this appliance may not correspond with colored markings identifying
the terminals in your plug, proceed as follows:
24
+
-
NOTICE
If the power supply is connected to 220-240 V with the range set to 100–
120 V, the instrument can be severely damaged.
Electrophoresis Power Supply EPS 3501 XL Operating Instructions 28-9639-69 AA