Omega HHLM112SD User Manual page 8

Light meter with data logging sd card
Table of Contents

Advertisement

4. Choose a decimal point or comma to represent the decimal division between
integers and fractions (for example, American-style 20.88 vs. European-style
20,88). Once you have finished setting the time, the display will show the word
"bASIC" above the phrase "dEC". Press the SET
(American style) or "Euro" appear in the upper display, as desired. Press the
REC/ENTER button to store the selection.
5. Enable or disable auto power off. Once the format of decimal divisions has
been set and stored, the display will show the word "yES' over the term "PoFF".
Press the SET
scheme ("yES" for enable; "no" for disable) is displayed. Press the REC/ENTER
button to store the selection. If enabled, the power off function shuts off the
DLM112SD after a period of inactivity of ten minutes
6. Enable or disable the beeper. Once the desired power management function
has been set and stored, the display will show the word "yES" over the word
"bEEP". Press the SET
displayed, and then press the REC/ENTER button to store the selection.
7. Select the thermocouple type. Once the beeper has been enabled or disabled,
the display will show the word "Set" over the word "tYPE". Press the SET
button until the desired setting ("K" or "J", for a Type K or Type J thermocouple)
appears on the right side of the display. Then press the REC/ENTER button to
store the selection.
8. Select the temperature unit. Once the thermocouple type has been set and
stored, the display will show the term "t-CF". Press the SET
desired setting ("F" or "C", for Fahrenheit or Celsius units) appears on the right
side of the display. Then press the REC/ENTER button to store the selection.
9. Set the datalogging sampling time. Once the temperature unit has been set
and stored, the display will show a value above the letters "SP-t". Press the SET
button or the LOGGER
desired sampling time appears above "SP-t". The options are 0, 1, 2, 5, 10, 30,
60, 120, 300, 600, 1800 and 3600 seconds (0 seconds to 1 hour).
10. Format the SD card. Once the sampling time has been set and stored,
the word "no" or "yES" will be displayed over the term "Sd-F". Press the SET
button to make a selection.
Choose "yES" whenever a new SD card is being used, or when a used card is being
repurposed (from use with another SD card instrument or a camera, for example) and
all data on it is to be erased. Choose "no" to preserve any data on a card previously
used with this instrument.
button until the desired automatic power off management
button until the desired setting ("yES" or "no") is
button to decrease or increase the value until the
button to make "bASIC"
button until the