SECTION 6—INDICATOR LIGHTS
SECTION 6—INDICATOR LIGHTS
INDICATOR LIGHT
COLOR
None
O
BELOW NORMAL -
2
YELLOW
FILLING - GREEN
FULL - GREEN
ATTENTION - RED
(Audible Alarm)
HomeFill
Compressor
®
COMPRESSOR
STATUS
Compressor is Off.
Compressor is On and
warming up
(0-3 minutes after
power is On).
O
from concentrator is
2
< 90% (any time after
three minute warmup).
Compressor is On and
cylinder is filling. O
from concentrator is
>90% (any time after
three minute warmup).
Compressor is On. Cyl-
inder is finished
filling (full) -remove
cylinder (FULL light is
On in >10 mins.)
or
Cylinder is not attached
properly - check con-
nection (FULL light is
On in <10 mins.).
Compressor is On.
Cylinder NOT filling
Causes:
Full cylinder attached
when unit turned On -
replace with empty
cylinder
or
Internal failure -
contact dealer.
30
EFFECT
O
flowing to user if
2
concentrator is on.
Cylinder filling not
started.
O
flowing to user if
2
concentrator is On.
Cylinder filling stops, will
restart when O
O
flowing to user if
2
concentrator is On.
Cylinder is filling.
2
O
flowing to user if
2
concentrator is On.
Cylinder filling stops.
Cannot fill another
cylinder until
compressor is turned Off
and On again.
O
flowing to user if
2
concentrator is On.
Compressor does not
begin to fill or stops
filling.
Unit must be turned Off
and On before
filling can occur.
O
flowing to user if
2
concentrator is On.
Part No 1100873
> 90%.
2

Advertisement