Select Mode & Setup
Operation
Press
Display Shows
9. If required,
enable (√)
to
or disable
(X) the
enable
functions
quick
edit
mode
10. Select
or
Period
from the
Setup
then
screen to
enable
measure-
ment
intervals to
be set
Page 30 - CEL-450/490 Operators Handbook
Comments
Use
and
to select a
field.
Use
and
to change
the entry in the field.
Save the settings by
pressing
again to
resume normal operation.
Then press
to return to
the Setup screen.
This screen allows a period
to be specified for regular
sequential measurement
and a shorter interval to be
set for the storage of
profiles.
Period measurement is
available only with a
CEL-490.
The predicted maximum run
time with the currently
selected parameters and
periods and the available
unused memory are also
indicated. Period
measurement is possible
only with a CEL-490.
The broadband measure-
ment period can be set to:
1, 5, 10, 15, 20, 30 seconds,
1, 5, 10 ,15, 20 , 30 minutes,
1 hour, - - - - (=none).
When available, the profile
interval must be equal to or
shorter than the period
setting.
The profile interval can be
set to:
10, 20, 50, 100, 250, 500,
milliseconds,
1, 5, 10, 15, 20, 30 seconds,
1, 5 10, 15, 20, 30 minutes.
Profiles for any of the
enabled parameters can be
stored.
Press
to access quick
edit to specify measurement
and profile periods.