1. Turn the therapy device off, turn the humidifier knob to 0, and allow approximately 15 minutes
for the heater plate and water to cool.
2. Open the humidifier door, and then slide the water tank out of the humidifier base. Empty any
remaining water.
3. Open the tank latch, and then gently pull apart the top and bottom halves of the humidifier tank
by pulling upward on the tank inlet or outlet. The inlet and outlet seals are removable.
4.
Wash the tank in a solution of warm water and a mild liquid dish detergent. Gently wash the
middle seal; it is a darker color for easy identification. Rinse the parts with clean water. Wipe the
parts completely on the top and bottom. Allow them to air dry.
M Series Heated Humidifier
S
ERVICE
To avoid electrical shock, always unplug the power cord from the wall outlet
before cleaning the humidifier.
LEANING
Empty and clean the water tank daily to prevent mold and bacteria growth.
Wipe the seal completely on the top and bottom sides.
Allow the water in the tank to cool to room temperature before removing the
tank from the humidifier.