Extech Instruments 382357 User Manual

Extech Instruments 382357 User Manual

Ground resistance clamp-on tester
Hide thumbs Also See for 382357:

Advertisement

Quick Links

USER MANUAL 
 
Ground Resistance Clamp‐On Tester 
MODEL 382357 

Advertisement

Table of Contents
loading

Summary of Contents for Extech Instruments 382357

  • Page 1 USER MANUAL    Ground Resistance Clamp‐On Tester  MODEL 382357 ...
  • Page 2: International Safety Symbols

    Introduction  Congratulations on your purchase of Extech’s 382357 Ground Resistance Tester. This Clamp‐on  device allows the user to measure ground resistance without the use of auxiliary ground rods.  This device can only be used on multi‐grounded systems. It is not necessary to disconnect the  ground under test. This meter is shipped fully tested and calibrated and, with proper use, will  provide years of reliable service. Please visit our website (www.extech.com) to check for the  latest version of this User Guide, Product Updates, and Customer Support.    Safety   Only qualified technicians should operate this meter.   Use extreme caution when operating the meter near energized electrical equipment.   Do not attempt to use this meter to twist or pry the ground electrode or ground wire away  from grounded equipment.   All metal objects or wires connected to the electrical system under test should be assumed  to be lethal until tested. Grounded systems are no exception.   Ensure that the batteries are inserted correctly in the battery compartment.    Remove the batteries from the meter if the meter is to be stored for long periods.  WARNING: If the meter is used in a manner not specified by the manufacturer, the built in  protections provided by the meter may be impaired.    International Safety Symbols  This symbol, adjacent to another symbol or terminal, indicates the user must  refer to the manual for further information.    This symbol, adjacent to a terminal, indicates that, under normal use, hazardous  voltages may be present    382357‐EU‐EN V2.4   2/13 ...
  • Page 3 Description  Jaw Assembly  Hold Button  Rotary Selector Switch  LCD Display  Record Button  ▼(decrement) Button  ▲(increment) Button  Function Button  Measurement Trigger          Display Symbols  REC: Indicates data logging in progress  NO. Indicates the READ function        :  Clamp jaws are not fully closed  NOISE: Excessive noise on the ground  conductor or rod  ))): Hi Lo alarm position  Function or Record display  Ω: Ohms (resistance measurement)  mA: milli‐amperes, Amperes (current measurements)  H:  Hold function activated  Main display           : Low Battery  Auto Power Off  382357‐EU‐EN V2.4   2/13 ...
  • Page 4: Operation

    Operation  Note: Ensure that the jaws are fully closed before testing.    Note: DO NOT clamp on to any conductor or open the clamp jaws during the start‐up self‐ calibration (self‐cal is identified by the LCD counting down from CAL7 to CAL1).  Ground Resistance Measurements  Open the clamp jaws and check that all surfaces are  clean and free of dust, dirt, or any foreign  substances. Clean if necessary.  Forcefully snap the clamp jaws open and closed  several times.  Turn the meter on by moving the rotary switch to   the ohms   position. DO NOT interrupt the start up  procedure by opening the clamp jaws or clamping  around a conductor. During self‐calibration, the  meter will display CAL5, CAL4, CAL3, CAL2 and  CAL1.  Once the start‐up calibration procedure is completed, the meter will sound a beep. Clamp  on to an electrode or ground rod and read the value of ground resistance on the display.  Ground / Leakage Current Measurement  Turn the meter on by moving the rotary switch to the  mA or A position.  Clamp on to the electrode or conductor  Read the value of leakage current displayed on the  LCD.        HOLD  Press the HOLD button to freeze the reading in the display. H will appear in the display. Press  HOLD again to exit the function.   382357‐EU‐EN V2.4   2/13 ...
  • Page 5: High And Low Alarm

    High and Low Alarm  Turn the meter on by moving the rotary switch to the ))) position.  Press the FUNC button to view and adjust the “HI” alarm value.  Press ▲ or the ▼ button to increase or decrease the value. Hold down the button to  increase the speed of adjustment. The value can be incremented from 0 to 1500 ohms and  then OL (overload). The value will roll over to 0 after OL is displayed.  Once the desired value is set, press the FUNC button a second time to view and adjust the  “LO” alarm value.  Press the FUNC button three more times to exit the setting mode.  The meter will now compare the measured resistance value with the high and low limits  when the meter is set to the ))) position. If the reading is larger than the HI value, the  meter will beep and display “HI” on the LCD. If the reading is lower than the LO value, the  meter will beep and display “LO” on the LCD.    Data Logging  Setting the Sampling Time  Press the FUNC button three times until the “SEC” symbol appears in the LCD display.  The sampling rate is now displayed in seconds.  Press the▲ or the ▼ button to increase or decrease the value. The value can be  incremented from 0 to 255 seconds.  Once the desired value is set, press the FUNC button several times until the characters in  the upper row of the LCD disappear.  Recording Data  Press the REC button to start recording data at the specified sampling rate. The “REC”  symbol will appear on the top of the LCD display. To stop recording, press the REC button.  The “REC” symbol will no longer be displayed on the LCD.  NOTE: Data logging will automatically stop if the memory is filled (116 records), or if the meter  detects a low battery condition.  NOTE: If the sampling interval is set to 0, only one sample will be recorded. To record another  sample, press REC again.  382357‐EU‐EN V2.4   2/13 ...
  • Page 6: Recalling Stored Data

    Recalling Stored Data  Press the FUNC button four times until the “NO.” symbol is displayed on the LCD. The  current record number will be displayed in the top row and the data will be displayed below  it.  Press the ▲ or the ▼ button to step through and read the data in the next memory  location. Holding down the button will increase the speed of the incrementing value.  Clearing Data Memory  With the meter powered off, press and hold down the REC button while turning the meter  on. The “CL” symbol will appear on the display indicating that memory has been cleared.  382357‐EU‐EN V2.4   2/13 ...
  • Page 7: Measurement Principles

       ,...           where:       If R  and R , R , R …R , are approximately the same, and n is large (200, for example), then R   will be much less then R  and may possibly approach zero.   Example:   If R  and R , R , R …R  are all 10  respectively and n = 200, then R  by calculation equals:                 382357‐EU‐EN V2.4   2/13 ...
  • Page 8 Turn the meter switch to the Ohm   function.  Once the start‐up calibration procedure is completed clamp on to an electrode or ground  rod.  Allow several seconds for the meter to settle and then note the reading.                Cellular Transmission Towers  WARNING: Use extreme care when making measurements around transmission towers; High  Voltage may be present.  The ground conductor should be located at the base of the tower, but your configuration may  be different.  Locate the ground conductor  Clamp around the ground conductor. You should place the clamp before any splices, bonds  or other splits in the ground system.  Read the measured reading.      Primary Ground Conductor Transmission   Tower       Cement Pad     Branch Ground Conductors 382357‐EU‐EN V2.4   2/13 ...
  • Page 9: Troubleshooting

    Service Entrance  Note: Multiple ground rods, multiple grounds i.e. ground rod(s) and water pipe grounds, or a  combination may be present depending on the particular situation. In these situations it is  necessary to make your measurements between the service‐entrance neutral and all  subsequent ground points.          Distribution   Panel     Water Pipe   Ground         Troubleshooting  Snap the jaws open and closed two or three times before turning the meter on  Do not clamp on to any device during the power‐up initialization sequence.   Snap the jaws two or three times after clamping on to the ground electrode or rod.  Slight amounts of drift may occur at high resistance measurements; this is normal and  should not be cause for concern.  Blank display or any failure mode  Replace the battery before proceeding.  Start‐up Initialization Failure  The start‐up calibration will continue indefinitely if the jaws are open or dirty. Check that the  jaws’ mating surfaces are clean. Do not open the jaws while the initialization sequence is  running.   Noise on the Ground Electrode or Rod  If noise is present (over 3A or 30V) on the ground electrode or rod, the word “NOISE“ will  appear on the display and readings can no longer be considered accurate. Before the meter can  be used accurately, the noise problem must be addressed.  382357‐EU‐EN V2.4   2/13 ...
  • Page 10 Reference Test Loop  The reference test loop is a set of resistors that provides a means to check the performance of  the meter.  Open the measurement jaws and check that the mating surfaces are clean and free of dust,  dirt, or any foreign substances. Clean if necessary.  Clamp the meter around the reference loop as shown in the diagram.  Give the meter a few seconds to settle and read the measurement. The meter should  display approximately the same value as marked on the reference loop, taking into  consideration the accuracy listed in the specification chart. If the readings are not similar to  what is marked on the reference loop, consult the Trouble Shooting and Maintenance  sections of this manual.                382357‐EU‐EN V2.4   2/13 ...
  • Page 11: Maintenance

    Maintenance  General Care  Store the ground resistance tester and reference loop in its case when not in use.  A damp  cloth may be used to clean any surface dirt from the meter. Never use harsh detergents,  abrasives, solvents, or cleaners on the meter.    Battery Replacement  When the low battery symbol appears on the display, replace the meter’s 9V battery.  Use a  high quality alkaline battery whenever replacement becomes necessary.  Remove the two rear screws and open the meter housing.  Replace the battery, close the meter housing, and affix the two rear screws.    Never dispose of used batteries or rechargeable batteries in household waste.  As consumers, users are legally required to take used batteries to appropriate  collection sites, the retail store where the batteries were purchased, or wherever  batteries are sold.  Disposal: Do not dispose of this instrument in household waste. The user is  obligated to take end‐of‐life devices to a designated collection point for the disposal  of electrical and electronic equipment.     Other Battery Safety Reminders  o Never dispose of batteries in a fire. Batteries may explode or leak.  o Never mix battery types. Always install new batteries of the same type.    Clamp Jaw Maintenance  The jaws may be brushed clean with a toothbrush or similar soft‐bristled brush.  Care must be  taken to ensure that the contact fins are not bent or deformed, as this will affect the operation  of the meter.  382357‐EU‐EN V2.4   2/13 ...
  • Page 12: General Specifications

    Datalogging sample rate  1 to 255 seconds  Power supply  One (1) 9V battery  Power consumption  40mADC  Battery Life  3000 measurements  Temp. Coefficient   0.15 times the specified accuracy per  C from 4 to 18 C (39 to  64ºF) and 28 to 50 C (82 to 122ºF)  Current Overload   Protected to 100A Continuous; 200A for less than 5 seconds  (50/60Hz)  Range selection  Automatic ranging  Safety  Meets the requirements for IEC1010‐1 Category III 300V and  Category II 600 V    Operating conditions  to 50 C (32 to 122 F) with < 85% RH    Storage conditions  ‐20 to 60 C (‐4 to 140 F) with < 75% RH  Dimensions  257 x 100 x 47mm (10.1 x 3.9 x 1.9”)  Weight  1.4lbs (640g)  Accessories  Resistance check plate, 9V battery, & carrying case  382357‐EU‐EN V2.4   2/13 ...
  • Page 13   0.300 to 1.000mA  ± (2.0% ± 0.05mA)  1.00 to 10.00mA  ± (2.0% ± 0.03mA)  Autoranging  10.0 to 100.0mA  ± (2.0% ± 0.3mA)  100 to 1000mA  ± (2.0% ± 0.3mA)  0.20 to 4.00A  ± (2.0% ± 0.03A)  Manual range  4.00 to 30.00A  ± (3.0% ± 0.03A)  Notes: 50/60Hz bandwidth, True RMS sensing, Crest Factor < 3.0    High and Low Alarm (programmable)    Range  Resolution    High Alarm  0‐1510       Low Alarm  0‐1510       Copyright © 2013 FLIR Systems, Inc.  All rights reserved including the right of reproduction in whole or in part in any form  ISO‐9001 Certified  www.extech.com    382357‐EU‐EN V2.4   2/13 ...

Table of Contents