SET UP

1.
Place batteries in the battery holder (see "Replacing Batteries" below).
2.
Plug the scale's AC adapter into the display assembly and into the power source.
3.
Press the ON/OFF button to turn the scale on. The display will show "start" and
then "0.0".
4.
When the self-test function is complete:
The digital display should read
"LB" should be indicated on the right side of the display. (To change the default
units of the scale to kilograms (Kg), please contact our Technical Support Center
at 1.800.638.3722 or 1.708.598.9100).
5.
Place a weight [not to exceed 600 Lb (270 Kg)] on the scale. The digital display will
------
blink "
" until the weight calculation is complete.
------
6.
Press the LBS/KG button to select the weight mode, either in kilograms or pounds.
7.
Press the REWEIGH button; the digital display should read again the weight of the item
on the scale.
8.
Remove the weight from the scale. The digital display should continue to read the
weight of the item that was on the scale.
9.
Press the ZERO button; the digital display should return to zero.
10.
Disconnect the power source connector from the scale, then turn the scale on using
ON/OFF button. The scale should automatically switch to battery power.
NOTE:
If the set up procedure failed, go to the troubleshooting instructions. If the problem is not
corrected, refer to qualified service personnel.
Replacing Batteries
1.
Disconnect the scale from the AC power source.
2.
Unscrew two plastic screws and carefully remove the battery holder from the display
assembly.
3.
Release attachable strap from the batteries. Replace the batteries with new ones and
reattach the strap.
4.
Position the battery holder into the display assembly and reinstall two plastic screws.
"0.0".
Figure 2. Replacing Batteries