This message/symbol
and means:
appears...

The test strip drum does not rotate.
Press the Test button to turn the
meter off. Open the drum compart-
ment and press the red drum release
button. Close the drum compartment
again.
150
All test strip drums from the same
pack have a bar code that is valid
for this pack only. Do not code the
meter with test strip drums from
other packs. If the test strips and
bar code do not belong together,
inaccurate blood glucose results
may be obtained. Test strip drums
from the same pack have the
same lot number. Compare the
lot numbers. These are printed on
the label of the test strip next to
the
symbol.

Symbols, display messages and troubleshooting
This message/symbol
and means:
appears...

Symbols, display messages and troubleshooting
The meter then rotates the drum
and reads the bar code. The drum
symbol and bottle symbol then flash
in the display. The 90-day use-by
period will no longer be properly
calculated for this drum. Perform a
quality control check. You may use
up all the test strips remaining in the
drum. The drum symbol continues
to flash in the display until you have
inserted a new drum.
The bar code on the test strip drum
is damaged. To troubleshoot, see er-
ror message E-2.
The small window covering the opti-
cal system is dirty.
Clean the window (see Chapter 7.1
"Cleaning the meter") and repeat the
test with a new test strip.