2. Preparation
2.1 Installing/Replacing the Batteries
1.
Remove the battery cover.
1) Push the hook on the bottom of
Wrist Blood
the battery cover.
Pressure Monitor
2) Pull the cover off the main unit.
Model HEM-6200
Instruction Manual
2.
Insert two 1.5V "AAA" alkaline (LR03)
batteries as indicated in the battery
5325545-5C
compartment and then replace the
battery cover.
Introduction
Thank you for purchasing the OMRON HEM-6200 Wrist Blood Pressure
Notes:
Monitor.
• If the battery low symbol (
The OMRON HEM-6200 is a compact and easy to use blood pressure monitor,
then replace both batteries at the same time.
operating on the oscillometric principle. It measures your blood pressure and
• The measurement values continue to be stored in memory even after
the batteries are replaced.
pulse rate simply and quickly. For comfortable controlled inflation without the
need of pressure pre-setting or re-inflation the devices uses its advanced
Disposal of used batteries should be carried out in accordance with the
"IntelliSense" technology.
national regulations for the disposal of batteries.
Please read this instruction manual thoroughly before using the unit.
Please keep for future reference.
2.2 Setting the Date and Time
For specific information about your own blood pressure, CONSULT YOUR
DOCTOR.
1.
Set the monitor to the correct date and time before
taking a measurement for the first time.

Important Safety Information

Press the
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia
or arteriosclerosis.
Please read this section carefully before using the unit.
Warning:
• Indicates a potentially hazardous situation which, if not avoided, could result
in death or serious injury.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and
self-treatment are dangerous.
• People with severe blood flow problems, or blood disorders, should consult a
doctor before using the unit, as cuff inflation can cause internal bleeding.
(Battery Usage)
• If battery fluid should get in your eyes, immediately rinse with plenty of clean
water. Consult a doctor immediately.
Caution:
• Indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate injury to the user or patient or damage to the equipment or
Year
other property.
(General Usage)
• Do not leave the unit unattended with infants or persons who cannot express
their consent.
• Do not use the unit for any purpose other than measuring blood pressure.
• Do not disassemble the unit or wrist cuff.
• Do not inflate the wrist cuff over 299 mmHg.
• Do not use a mobile phone or other devices that emit electromagnetic fields,
2.
Press the START/STOP button to store the setting.
near the unit. This may result in incorrect operation of the unit.
• Do not operate unit in a moving vehicle (car, airplane).
Notes:
(Battery Usage)
• To reset the date and time, hold down the
• If battery fluid should get on your skin or clothing, immediately rinse with
is off.
plenty of clean water.
• If the batteries have been removed for 30 seconds or more, the date
• Use only two "AAA" alkaline (LR03) batteries with this unit. Do not use other
types of batteries.
and time setting will need to be reset.
• Do not insert the batteries with their polarities incorrectly aligned.
• If the date and time are not set, "-:--" appears during or after
• Replace old batteries with new ones immediately. Replace both batteries at
measurement.
the same time.
• Remove the batteries if the unit will not be used for three months or more.
• Do not use new and used batteries together.

3. Using the Unit

General Precautions
• Do not apply strong shocks and vibrations to or drop the unit.

3.1 Applying the Wrist Cuff

• Do not take measurements after bathing, drinking alcohol, smoking,
exercising or eating.
Do not apply over clothing.
• Do not inflate the wrist cuff when it is not wrapped around your wrist.
• Do not wash the wrist cuff or immerse it in water.
• Read and follow the "Important information regarding Electro Magnetic
1.
Place the wrist cuff over your wrist.
Compatibility (EMC)" in the Technical Data Section.
• Read and follow the "Correct Disposal of This Product" in the Technical Data
Section when disposing of the device and any used accessories or optional
parts.
1. Overview

Main unit:

2.
Wrap the wrist cuff around your wrist.
A
Your thumb should face upward.
C
B
D
E
F
G
Make sure that the wrist cuff does not cover
the protruding part of the wrist bone (ulna) on
A. Display
E. START/STOP button
the outside of the wrist.
B. Battery compartment cover
F. SET (
) button
C. Wrist cuff
G. Date/Time setting (
) button
D. MEM (Memory) button
Notes:
Display:
• You can take a measurement on either
H
I
your left or right wrist.
J
K
L
• Wrap the wrist cuff securely around the
M
wrist for taking accurate measurements.
N
O
S
R
Q
P
H. Systolic blood pressure
O. Battery low symbol
I. Diastolic blood pressure
• The blood pressure can differ between the right
P. Date/Time display
arm and the left arm, and therefore also the
J. Memory symbol
Q. Heartbeat symbol
K. Average value symbol
measured blood pressure values can be
(Flashes during measurement.)
different. Omron recommends to always use
L. Irregular heartbeat symbol
R. Deflation symbol
the same arm for measurement. If the values
M. Pulse display
S. Blood pressure level indicator
N. Movement error symbol
between the two arms differ substantially,
please check with your doctor which arm to use
for your measurement.

3.2 How to Sit Correctly

To take a measurement, you need to be relaxed and comfortably seated, under
comfortable room temperature. No eating, smoking or exercising 30 minutes
before taking a measurement.
• Sit on a chair with your feet flat on the floor.
2
Sit upright with your
back straight.
1
) appears on the display, turn the unit off
Relax your wrist and hand.
button.
Do not bend
Do not clench
1) Push the
button to change.
your wrist back.
your fist.
• Hold down to advance the
digit rapidly.

3.3 Taking a Reading

2) Push the
button to confirm.
Notes:
Next setting appears.
• To cancel a measurement, press the START/STOP button at any time
during measurement.
• Remain still while taking a measurement.
1.
Press the START/STOP button.
The wrist cuff will start to inflate automatically.
START
INFLATE
Month
Day
Hour
Minute
button while the power
2.
Undo the wrist cuff and remove the unit.
3.
Press the START/STOP button to turn off the monitor.
The monitor automatically stores the measurement in its memory.
It will automatically turn off after two minutes.
Important:
• Recent research suggests that the following
values can be used as a guide to high blood
pressure for measurements taken at home.
Systolic Blood Pressure
Above 135 mmHg
Diastolic Blood Pressure Above 85 mmHg
This criteria is for home blood pressure
measurement.
• Your blood pressure monitor includes an irregular
heartbeat feature. Irregular heartbeats can influence
the results of the measurement. The irregular
heartbeat algorithm automatically determines if the
measurement is usable or needs to be repeated. If
the measurement results are affected by irregular
heartbeats but the result is valid, the result is shown
together with the irregular heartbeat symbol (
the irregular heartbeats cause the measurement to
be invalid, no result is shown. If the irregular
heartbeat symbol (
) is shown after you have taken a measurement,
repeat the measurement. If the irregular heartbeat symbol (
shown frequently, please make your doctor aware of it.

3.4 Using the Memory Function

ulna
The monitor automatically stores the result up to 60 sets. It can also
calculate an average reading based on the measurements from the last
three readings taken within 10 minutes. If there are only two readings in
memory for that period, the average will be based on two readings. If
there is one reading in memory for that period, the average will be
based on one reading.
Notes:
• If the memory is full, the monitor will delete the oldest readings.
• When viewing the reading taken without setting the date and time,
"-:--" is displayed instead of the date and time.
Average Value
View the Average
View Previous Reading
Value
• The Memory number
• If the previous reading
appears for a second
was taken without
before the pulse rate is
setting the date and
displayed. The newest
time, the average
set is numbered "1",
value is not calculated.
To delete all the values stored in memory
When the memory symbol (
) appears, first press the MEM button. Then
while holding it down, press the START/STOP button simultaneously for
about 2-3 seconds.
The cuff should be at the
First
Second
same level as your heart.
Note: You cannot partially delete the stored readings.

4. Troubleshooting and Maintenance

4.1 The Icons and Error Messages
Error Display
Cause
Remedy
Remove the wrist cuff. Wait
2-3 minutes and then take
Do not use your other hand
Irregular or weak pulses
another measurement.
to support the wrist cuff.
are detected.
Repeat the steps in section
This can result in inaccurate
3.3. If this error continues to
measurement results.
appear, contact your doctor.
Movement during
Carefully read and repeat the
measurement.
steps in section 3.3.
You should replace them with
new ones ahead of time.
Blink
The batteries are low.
Refer to section 2.1.
Do not bend
You should replace them with
your wrist forward.
The batteries are
new ones at once.
Lit
exhausted.
Refer to section 2.1.
Cuff is over inflated.
Carefully read and repeat the
steps listed under section 3.3.
Movement during
measurement.
DEFLATE
COMPLETE
An Er mark with a
code/number indicates
Consult your OMRON
the device has a
retail outlet or distributor.
hardware failure.
Alternating
date/time
display

4.2 Troubleshooting

Problem
Cause
Remedy
Measure while in the
The wrist cuff is not at
correct posture. Refer
heart level.
to section 3.2.
The cuff is wrapped
Wrap the cuff correctly.
snugly around the
Refer to section 3.1.
wrist.
The reading is
extremely low (or
Relax and try taking the
Your arms and
high).
measurement again.
shoulders are tense.
Refer to section 3.3.
Remain still and do not
Movement or talking
talk during
during measurement.
measurement. Refer to
section 3.3.
Consult your OMRON
Wrist cuff pressure
Air is leaking from the
retail outlet or
does not rise.
wrist cuff.
distributor.
Apply the cuff correctly
so that it is securely
Wrist cuff deflates too
The wrist cuff is
). If
wrapped around the
soon.
loose.
wrist. Refer to section
3.1.
Blood pressure readings
) is
constantly vary with time
of day and how relaxed
The blood pressure is different each time.
you are. Take several
The reading is extremely low (or high).
deep breaths and try to
remain relaxed before
taking a measurement.
The unit loses power
The batteries are
Replace the batteries
during measurement.
drained.
with new ones.
The batteries are
Replace the batteries
drained.
with new ones.
Nothing happens
when you press the
The batteries have
Insert the batteries with
buttons.
been inserted
the correct (+/-) polarity.
incorrectly.
• Press the START/STOP button and repeat
Previous Reading
measurement.
• If the problem continues, try replacing the
Other problems.
batteries with new ones.
If this still does not solve the problem, contact
your OMRON retail outlet or distributor.
Alternating
date/time display
View Readings Stored
in Memory
• Hold down to view
readings rapidly.
4.3 Maintenance
Important information regarding Electro Magnetic Compatibility (EMC)
To protect your unit from damage, please avoid the following:
With the increased number of electronic devices such as PC's and mobile
• Subjecting your unit to extreme temperatures, humidity,
(cellular) telephones, medical devices in use may be susceptible to
or direct sunlight.
electromagnetic interference from other devices.
Electromagnetic interference may result in incorrect operation of the medical
• Washing the cuff or exposing the cuff or unit to water.
device and create a potentially unsafe situation.
• Disassembling the unit.
Medical devices should also not interfere with other devices.
• Subjecting the unit to strong shocks or vibrations.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility)
Dropping the Unit.
with the aim to prevent unsafe product situations, the EN60601-1-2:2007
• Cleaning the unit with volatile liquids. THE UNIT
standard has been implemented. This standard defines the levels of immunity to
SHOULD BE CLEANED WITH A SOFT, DRY CLOTH.
electromagnetic interferences as well as maximum levels of electromagnetic
emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this
EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate
strong electrical or electromagnetic fields, near the medical device. This may
result in incorrect operation of the unit and create a potentially unsafe situation.
Recommendation is to keep a minimum distance of 7 m. Verify correct
operation of the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is available at OMRON
HEALTHCARE EUROPE at the address mentioned in this instruction manual.
Documentation is also available at www.omron-healthcare.com.
Correct Disposal of This Product
Use a soft, moistened cloth and soap to clean the cuff.
(Waste Electrical & Electronic Equipment)
Keep the unit in its storage case when not in use.
This marking shown on the product or its literature, indicates that it
Fold the cuff into the storage case.
should not be disposed of, with other household wastes at the end
Do not store the unit in the following situations:
of its working life. To prevent possible harm to the environment or
• If the unit is wet.
human health from uncontrolled waste disposal, please separate
• Locations exposed to extreme temperatures, humidity,
this from other types of wastes and recycle it responsibly to
promote the sustainable reuse of material resources.
direct sunlight, dust or corrosive vapours.
• Locations exposed to vibrations, shocks or where it will
Household users should contact either the retailer where they purchased this
product, or their local government office, for details of where and how they can
be at risk of falling.
take this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions
Calibration and Service
of the purchase contract. This product should not be mixed with other commercial
• The accuracy of this blood pressure monitor has been
wastes for disposal.
carefully tested and is designed for a long service life.
This product does not contain any hazardous substances.
• It is generally recommended to have the unit inspected
every two years to ensure correct functioning and
accuracy. Please consult your local authorised OMRON
distributor or dealer.

5. Technical Data

Product Description
Wrist Blood Pressure Monitor
Asia Pacific HQ
Model
HEM-6200
Display
LCD Digital Display
Measurement Method
Oscillometric method
Measurement Range
Pressure: 0 mmHg to 299 mmHg
Pulse: 40 to 180 beats/min.
Accuracy
Pressure: ±3 mmHg
Pulse: ±5% of reading
Inflation
Automatic inflation by pump
Deflation
Automatic rapid deflation
Memory
60 Measurements
Power Source
2 "AAA" alkaline (LR03) batteries 1.5V
Battery Life
Approx. 300 measurements with new alkaline
batteries at a room temperature of 23°C
Applied Part
= Type B
Protection Against
Internally powered ME equipment
Electric Shock
Operating temperature/
+10°C to +40°C / Maximum: 30 to 85% RH
Humidity
Storage temperature/
-20°C to +60°C / Maximum: 10 to 95% RH /
Humidity/ Air pressure
700-1060 hPa
Console Weight
Approximately 117g without batteries
Outer Dimensions
Approximately 71 (w) mm × 41 (h) mm × 70 (d) mm
without the wrist cuff
Measurable
Approximately 13.5 to 21.5 cm
circumference of wrist
Cuff Material
Nylon and polyester
Package Content
Main unit, storage case, two "AAA" alkaline (LR03)
batteries, instruction manual
Notes:
• Subject to technical modification without prior notice.
• Disposal of this product and used batteries should be carried out in
accordance with the national regulations for the disposal of electronic
products.
Please read this instruction manual carefully before using the device.
• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device
Directive).
• This blood pressure monitor is designed according to the European Standard
EN1060, Noninvasive sphygmomanometers Part 1: General Requirements
and Part 3: Supplementary requirements for electromechanical blood
pressure measuring systems.
HEM-6200-C1_B_M02_100309.pdf
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto
617-0002 JAPAN
OMRON HEALTHCARE EUROPE B.V.
Scorplus 33, 2132 LR Hoofddorp
THE NETHERLANDS
OMRON HEALTHCARE SINGAPORE PTE LTD.
438A Alexandra Road, #05-05/08
Alexandra Technopark, Singapore 119967
www.omron-healthcare.com.sg

Advertisement

loading

Summary of Contents for Omron HEM-6200

  • Page 1: How To Sit Correctly

    ) appears on the display, turn the unit off result in incorrect operation of the unit and create a potentially unsafe situation. The OMRON HEM-6200 is a compact and easy to use blood pressure monitor, then replace both batteries at the same time.
  • Page 2 (+/-) 617-0002 JAPAN OMRON HEALTHCARE EUROPE B.V. • START/STOP Scorplus 33, 2132 LR Hoofddorp • THE NETHERLANDS • OMRON HEALTHCARE SINGAPORE PTE LTD. 438A Alexandra Road, #05-05/08 Asia Pacific HQ OMRON Alexandra Technopark, Singapore 119967 www.omron-healthcare.com.sg • Omron • •...