  • Page 1 Dome Ex Display ® User’s Guide...
  • Page 2 NDSsi. NDS Surgical Imaging, NDSsi, the NDSsi logo, Dome, the Dome by NDSsi logo, Dome E2, Dome E2cHB, Dome E3, Dome E3cHB, Dome E3cHB DisplayPort, Dome E5, Dome CXtra, RightLight are either registered or unregistered trademarks of NDS Surgical Imaging.
  • Page 3: About The Display

    Bundled with the display system, the Dome CXtra software enhances the functionality of the Dome Ex display with a range of value-added services, such as DICOM calibration, error reporting, and backlight saver. Network management of Dome displays running the Dome CXtra software requires the Enterprise Management Service for Dome CXtra.
  • Page 4 DC power adapter* Power cords Display driver and DVI cable DisplayPort cable** calibration software * Some Dome Ex displays ship with a DC adapter that operates per power switch. ** For Dome E3cHB DisplayPort unit only. 2 | Dome Ex Display...
  • Page 5: System Requirements

    System Requirements All display systems require operating system Windows 7 (x86 or x64) or Windows XP Service Pack 3 (x86 or x64). IMPORTANT! Note the following: • NVIDIA boards. Display configurations using the NVS295 board require a DisplayPort-to-DVI adapter. Display configurations, except for the Dome E3cHB DisplayPort unit, using Quadro 600, Quadro 2000, or Quadro 4000 require a DisplayPort-to-DVI adapter for the second display.
  • Page 6: Display Components

    8-pin port on Dome E3cHB, and 6-pin port on Dome E3cHB DisplayPort. • Video input. Drives the data to the display. All Dome Ex displays have a standard DVI connector. The Dome E3cHB DisplayPort unit also includes a DisplayPort connector port.
  • Page 7: Installing The Display

    ‡ Refer to Dome E4c Display User’s Guide for instructions. Warning In locations where 240V outlets are used, connect the Dome Ex display only to a center-tapped, 240V, single-phase supply (for Canada and the United States only). If you leave the computer turned on, you could get an electric shock and cause damage to the system components.
  • Page 8: Installing The Display Controller

    5 For Quadro 4000 board only: Connect the 6-pin auxiliary power supply cable to the controller. (Use the built-in 6-pin connector if it is present.) 6 Secure the mounting bracket. 7 Reattach the computer cover and the peripherals. 6 | Dome Ex Display...
  • Page 9 Connecting the Video and Power Cables Turn your computer off. Leave the power cord plugged into the grounded outlet. Use the Dome Ex display with the power adapter and video cable shipped. IMPORTANT! Note the following: •...
  • Page 10: Installing The Display Driver

    5 Follow the onscreen instructions to complete the installation. When the Setup complete message appears, select Yes, I want to restart my computer now. Click Finish. 6 If the Digital Signature Not Found message appears, click Yes or Continue to complete the driver installation. 8 | Dome Ex Display...
  • Page 11: Adjusting The Display Properties

    Adjusting the Display Properties After installation, use the Display Properties dialog to makes changes to the video settings or to set up a multiple display configuration. For Windows 7 To set display properties 1 Right-click the desktop and select Personalize. 2 Click Display on the left panel of the dialog.
  • Page 12 2 Select the monitor that you want to change. Then click Advanced. 3 Click the Monitor tab. Make sure the check box for Hide modes that this monitor cannot display is empty (unchecked). Click OK. 10 | Dome Ex Display...
  • Page 13: Setting Brightness

    4 Click the Adapter tab, then click the List All Modes button. 5 Choose the mode and click OK. Click OK to finish. Display Resolution Orientation Dome E2 1200 x 1600 Portrait Dome E2cHB 1600 x 1200 Landscape Dome E3 1536 x 2048 Portrait Dome E3cHB...
  • Page 14: Power Management

    Power Management You have two ways to lower energy usage when the Dome Ex display is idle: • Dome CXtra Backlight Saver service (preferred) • Screen Saver (set via Windows operating system) Activate the power saver when you anticipate periods of inactivity, such as at the end of the work day.
  • Page 15: Technical Information

    Technical Information The design of the Dome® Ex digital display takes into account every known measure to ensure your personal safety. Improper use of the display can result in electric shock, fire, or damage to the display. Read all instructions before setting up the display. Classification: Shock Protection: Class I.
  • Page 16: Symbol Explanations

    TEMPERATURE. Transport and storage -10° to 60° C. Intended use The Dome Ex display is an AMLCD display designed for viewing medical X-ray images. This unit should not be used near patients and should be kept outside of 1.83 m perimeter and 2.29 m vertical.
  • Page 17: Safety Precautions

    • Keep the display dry if it is part of a surgical system. The display lacks protection against liquids or spills. • In locations where 240V outlets are used, connect the Dome Ex display only to a center-tapped, 240V, single-phase supply (for Canada and the United States).
  • Page 18: Unpacking And Handling Tips

    Unpacking and handling tips The Dome Ex display is a precision instrument that requires proper care to maintain product operation and adherence to specification. Unpack the display and components carefully, then set up and handle the unit properly to avoid damage to the LCD panel.
  • Page 19: Cleaning The Display

    Cleaning the display Observe the following guidelines to maintain the display and the LCD screen. • Use a clean, lint-free, absorbent cotton cloth to clear off any residual glue from removal of the protective film or to remove surface dust. Apply light pressure to remove the dust.
  • Page 20 Display weight (without stand) 5.5 kg (12 lb) Mounting options Desktop stand (standard); 100-mm VESA mounting (optional) Power Adapter, 4-pin power connector Ault MW116KA1249F53 (CAUTION: Use only the power adapter supplied with the display unit.) 18 | Dome Ex Display...
  • Page 21: Power Supply

    Power Supply CAUTION: Use only the power adapter supplied with the Dome E2 display unit; Ault MW116KA1249F53. Category Characteristic Item Specification Power input Voltage selection Auto-ranging requirements Voltage 100–240V AC Current 1.0 A Frequency 50–60 Hz ± Power output Voltage 12 V DC requirements Current...
  • Page 22 Display weight (without stand) 5.5 kg (12 lb) Mounting options Desktop stand (standard); 100-mm VESA mounting (optional) Power Adapter, 8-pin power connector BridgePower BPM130S12F02 (with power switch) (CAUTION: Use only the adapter supplied with unit.) 20 | Dome Ex Display...
  • Page 23 Power Supply CAUTION: Use only the power adapter supplied with the Dome E2cHB display unit; BridgePower BPM130S12F02 (130W). Category Characteristic Item Specification Power input Voltage selection Auto-ranging requirements Voltage 100–240V AC Current 3.0 A Frequency 50–60 Hz Power output Voltage 12 V DC ±5% requirements Current...
  • Page 24 Mounting options Desktop stand with integrated USB (standard); 100-mm VESA mounting (optional) Power Adapter, 8-pin power connector Ault MW122RA1223F52 or Ault MW122KA1223F52 BridgePower BPM130S12F02 (with power switch) (CAUTION: Use only the adapter supplied with unit.) 22 | Dome Ex Display...
  • Page 25 Power Supply CAUTION: Use only the power adapter supplied with the Dome E3 display unit; Ault MW122RA1223F52 or Ault MW122KA1223F52 (both 120W), or BridgePower BPM130S12F02 (130W). Category Characteristic Item Specification Ault MW122RA1223F52 or Ault MW122KA1223F52 Power input requirements Voltage selection Auto-ranging Voltage 100–240V AC...
  • Page 26 100-mm VESA compliant mounting; Desktop stand with integrated USB (standard) Power Adapter, 8-pin power connector JEC Korea Corporation, Type JMW1150KA2423F51 Adapter, 6-pin power connector JEC Korea Corporation, Type JMW1150KA2400F52 (CAUTION: Use only the adapter supplied with unit.) 24 | Dome Ex Display...
  • Page 27 Power Supply CAUTION: Use only the power adapter supplied with the Dome E3cHB display unit. Use JEC Korea Corporation, Type JMW1150KA2423F51 for displays with an 8-pin power input connector. Use JEC Korea Corporation, Type JMW1150KA2400F52 for displays with a 6-pin power input connector. Category Characteristic Item Specification...
  • Page 28 Display weight (with stand) 10.2 kg (22.5 lb) typical Mounting options 100-mm VESA compliant mounting; Desktop stand with integrated USB (standard) Power Adapter JEC Korea Corporation, Type JMW1150KA2400F52 (CAUTION: Use only the adapter supplied with unit.) 26 | Dome Ex Display...
  • Page 29 Power Supply CAUTION: Use only the power adapter supplied with the Dome E3cHB DisplayPort unit. Use JEC Korea Corporation, Type JMW1150KA2400F52. Category Characteristic Item Specification Power input Voltage selection Auto-ranging requirements Voltage 100–240V AC Current 3.0 A Frequency 50–60 Hz ±...
  • Page 30 10.2 kg (22.5 lb) with stand Mounting options Desktop stand (standard); 100-mm VESA mounting (optional) Power Adapter Ault MW122RA1223F52 or Ault MW122KA1223F52 BridgePower BPM130S12F02 (with power switch) (CAUTION: Use only the adapter supplied with unit.) 28 | Dome Ex Display...
  • Page 31 Power Supply CAUTION: Use only the power adapter supplied with the Dome E5 display unit; Ault MW122RA1223F52 or Ault MW122KA1223F52 (both 120W), or BridgePower BPM130S12F02 (130W). Category Characteristic Item Specification Ault MW122RA1223F52 or Ault MW122KA1223F52 Power input requirements Voltage selection Auto-ranging Voltage 100–240V AC...
  • Page 32: Video Modes

    Static gray Dome E3 1536 x 2048 Portrait True color Dome E3cHB 2048 x 1536 Landscape Dome E3cHB DisplayPort Dome E5 2048 x 2560 Portrait Dynamic gray 2560 x 2048 Landscape Nonlinear gray Static gray 30 | Dome Ex Display...
  • Page 33: Connector Ports

    Connector Ports All Dome Ex displays have a standard DVI connector. The Dome E3cHB DisplayPort unit also has a DisplayPort connector. Video Power Display Adapter Port Connector Input Dome E2 4-pin 12V ±5% Ault MW116KA1249F53 (80 W) Dome E2cHB 8-pin 12V ±5%...
  • Page 34 Quadro FX 380 2 Dual Link DVI-I Quadro 600 1 DisplayPort 1 Dual Link DVI-I Quadro 2000 2 DisplayPort 1 Dual Link DVI-I Quadro 2000D 2 Dual Link DVI-I Quadro 4000 2 DisplayPort 1 Dual Link DVI-I 32 | Dome Ex Display...
  • Page 35: Led Status Lights

    LED Status Lights The two LEDs on the back panel of the Dome Ex displays provide information about the display. • LED A describes the digital-link status between the display controller and the interface. • LED B describes the display-panel status. it shows any faults currently in the panel.
  • Page 36 Flashing green Fault Solid yellow DDC power-on, 12V/24V power not detected Alternating green/yellow POST failure NOTE: A red LED or combination of red LEDs indicates a hardware or software fault. Contact Dome Technical Support for assistance. 34 | Dome Ex Display...
  • Page 37: Troubleshooting

    Unable to set the The function Hide all When using commodity graphics maximum or native modes that this monitor boards with Dome Ex displays, make resolution of the cannot display is selected sure the check box for the function is display on a subdialog of the empty (unchecked).
  • Page 38: Regulatory Compliance

    (This applies to all Dome Ex displays except the Dome E3cHB, Dome E3cHB DisplayPort, and Dome E5 displays, which are Class A.) It is the system integrator’s responsibility to test and ensure that the entire system complies with applicable electromagnetic compatibility (EMC) laws.
  • Page 39 We offer state-of-the-art displays suitable to the European market, CE-marked displays based on compliance with counsel directive 93/42/EEC—commonly referred to as the Medical Device Directive (MDD). The following summarizes our qualification of these displays as it relates to compliance with the MDD. The European Medical Device Directive requires that the intended use of the device be defined.
  • Page 40 Surgical Imaging International Electrotechnical Commission (IEC) declaration of conformity. NDS Surgical Imaging, located in San Jose, Calif., USA, is the manufacturer of these displays in the meaning of the directive. As required by the MDD in Article 14, NDS Surgical Imaging, not residing in the European Economic Area (EEA), has a European representative, NDS Surgical Imaging —...
  • Page 41 Declaration of Conformity document, contact NDS Surgical Imaging. The Dome Ex digital flat-panel display meets the essential health and safety requirements, is in conformity with, and the CE marking has been applied according to the relevant EU Directives listed below, using the relevant section of the following EU standards and other normative documents;...
  • Page 42 U.S. FCC Compliance Statement Class B Digital Device: Dome E2, Dome E2cHB, or Dome E3 Display This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
  • Page 43 Class A Digital Device: Dome E3cHB, Dome E3cHB DisplayPort, or Dome E5 Display This device complies with Part 15 of the FCC Rules. NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment.
  • Page 44 Each Dome Ex display has one or more of the following certifications. China China Compulsory Certification regulating safety and EMC. GB4943-2001 GB9254-2008 CB17625.1-2003 Class A Dome E3cHB, Dome E3cHB DisplayPort, and Dome E5 Germany TÜV Rheinland GM Mark for medical products.
  • Page 45 Standard 3-year “repair and return” warranty * If, within 10 years of initial purchase, the maximum output of the Dome Ex/Sx grayscale display is determined by NDSsi to be less than 300 cd/m², NDSsi will repair or replace the display at its sole discretion.
  • Page 46 EXCLUSIONS – WHAT IS NOT COVERED 1 Any product with a defaced, modified, or removed serial number. 2 Damage, deterioration or a malfunction resulting from accident, misuse, neglect, fire, water, lightning, or other acts of nature, unauthorized product modification, or failure to follow instructions supplied with the product.
  • Page 47 NDSSI’S LIABILITY IS LIMITED TO THE COST OF REPAIR OR REPLACEMENT OF THE PRODUCT. NDSSI SHALL NOT BE LIABLE FOR THE FOLLOWING: 1 DAMAGE TO OTHER PROPERTY CAUSED BY ANY DEFECTS IN THE PRODUCT, DAMAGES BASED UPON INCONVENIENCE, LOSS OF USE OF THE PRODUCT, LOSS OF TIME, LOSS OF PROFITS, LOSS OF BUSINESS OPPORTUNITY, LOSS OF GOODWILL, INTERFERENCE WITH BUSINESS RELATIONSHIPS, OR OTHER COMMERCIAL LOSS, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGES.
  • Page 48 CONTACTING NDSSI CUSTOMER CARE PRODUCT LINE Dome® TOLL FREE (866) 961-9340 LOCAL (503) 620-3787 E-MAIL medicalsupport@ndssi.com HOURS . – 5 . Mon –Fri Pacific Time 1 Contact NDSsi Medical Support during the hours listed above. Have your NDSsi model number and serial number available. You may be required to provide proof of purchase as a condition for receiving warranty service.
  • Page 49 PAYMENT FOR NONWARRANTY EVALUATIONS AND REPAIRS AND OTHER EVALUATION CHARGES: Nonwarranty repairs requested by customer will be performed at NDSsi’s then current rates for any nonwarranty repairs and will include applicable evaluation fees and the cost of return shipping (if customer requests that NDSsi ship the repaired product).
  • Page 50 NDS Surgical Imaging, LLC 5750 Hellyer Avenue San Jose CA 95138-1000 USA www.ndssi.com...

Table of Contents