FLASH CODES

Flash codes indicate the nature of an abnormal condition
directly from the SHARK Information Gauge. Without the use of
any servicing tools, the condition can be simply diagnosed.
Flash
Description
Code
1
User Fault
2
Battery Fault
3
Left Motor Fault
4
Right Motor Fault
5
Left Park Brake Fault
6
Right Park Brake Fault
7
SHARK Remote Fault
8
SHARK Power Module
Fault
9
SHARK
Communications
Fault
10
Unknown Fault
11
Incompatible Remote
Frame
3 Years
Possible stall timeout or user error.
Release the joystick to neutral and try again.
Try charging the batteries.
Batteries may require replacing.
Check the batteries and cabling.
Check the left motor, connections and cabling.
Check the right motor, connections and cabling.
Check the left park brake, connections and cabling.
Check the right park brake, connections and cabling.
Check the SHARK Communications Bus connections and wiring.
Replace the Remote.
Check SHARK connections and wiring.
Replace the Power Module.
Check Battery voltage is greater than 17V.
Check SHARK Bus Cable.
Replace the SHARK Power Module.
Replace the SHARK Remote.
Check all connections and wiring.
Consult a service agent.
The Remote is incompatible with the Power Module.
Ensure the brand of the power Module matches that of the remote.
WARRANTIES
Electrical
Parts
1 Year
6 Months
Motor/Transaxle/Brake
1Year

Advertisement