Section 4 - Specifications - Invacare Platinum IRC5LXO2 Operator's Manual

Platinum series xl, 5, 10 oxygen concentrators standard, with senso2 and homefill ii
Hide thumbs Also See for Platinum IRC5LXO2:
Table of Contents

Advertisement

Available languages

Available languages

SECTION 4 —SPECIFICATIONS
SECTION 4 — SPECIFICATIONS
Electrical Requirements:
Rated Current Input:
Sound Level:
Altitude:
*Oxygen Output
Concentration Levels:
Maximum Outlet Pressure:
*NOTE: Stated concentration levels achieved after initial warm-up
period (approximately 30 minutes).
Platinum™Series
120 VAC ± 10% (132 VAC/108 VAC), 60
Hz
Platinum 5 and XL: 4.3 A
Platinum 10: 4.7 A
Platinum 5 and XL: 50 dB Average
Platinum 10: 58 dB Average
Platinum 5 and XL - Up to 6,000 ft. (1828
meters) above sea level without degradation
of concentration levels. From 6,000 ft. (1828
meters) to 13,129 ft. (4000 meters) below
90% efficiency
Platinum 10 - Up to 4,000 ft. (1230 meters)
above sea level without degradation of
concentration levels. From 4,000 ft. (1230
meters) to 13,129 ft. (4000 meters) below
90% efficiency
All 5LXO2/5LX models 95.6% to 87% at 0.5
to 5 L/min. and all 10LXO2/10LX models
95.6% to 87% at 2 L to 10 L/min.
Platinum 5 and XL: 5 psi ± 0.5 psi. (34.5 kPa
± 3.45 kPa.)
Platinum 10: 9 psi ± 0.5 psi. (62.0 kPa ± 3.45
kPa.)
14
Part No 1145728

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Platinum IRC5LXO2 and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

Platinum irc10lxo2Platinum irc10lxPlatinum irc5lx

Table of Contents