Monitor
Report
Condition
EC82
Kinked hose.
EC03
EC03
Patient canceled reading by pressing STOP key. No retry attempt is made
following an EC03 code.
EC13
The Office Check Mode has been reinstated. No retry attempt is made
following an EC13 code.
EC04
EC04
Blood pressure measurement not completed in the maximum time
allowed. Occasional EC04 messages may result from excessive patient
movement. Frequent EC04 messages would indicate an improperly
applied cuff or a monitor malfunction which requires service.
ECn4
(where n = 1 to 9) Indicates that one or more of the blood pressure results
have been corrupted and subsequently recovered. Frequent occurrence
of this message would indicate a malfunction which requires service.
EC05
The individual blood pressure result has been corrupted and cannot be
recovered.
EC15
Equipment malfunction. Return to Spacelabs Medical for service.
EC25
Unit failed to initialize. Please initialize.
90207
At least one of the blood pressures or time readings obtained before the
EC35
event code is erroneous. Interpret all readings with caution.
90217
The monitor needs to be reinitialized.
EC05
EC05 &
EC45
Invalid bleed size. The monitor automatically has changed the bleed size
EC45
to 8 mmHg.
EC05 &
EC55
An unexpected loss of power possibly caused by a) removal of the
EC55
batteries during a blood pressure measurement, b) hardware
overpressure, or c) a hardware time-out. Frequent EC55 messages would
indicate a malfunction which requires service.
EC05 &
EC65
90207
EC65
Equipment malfunction. Return to Spacelabs Medical for service.
90217
Extremely large artifact.
EC05 &
EC75
Equipment malfunction. Return to Spacelabs Medical for service.
EC75
EC05 &
EC85
Equipment malfunction. Return to Spacelabs Medical for service.
EC85
EC05 &
EC95
Cuff pressure baseline out of bounds. The monitor should correct the
EC95
baseline automatically within 10 minutes; or it can be set by initialization
of the monitor. If initialization does not correct the condition the monitor
must be returned to Spacelabs Medical for calibration.
Operation
1-15