Spacelabs 90207 Operation Manual page 13

Ambulatory blood pressure monitors
Hide thumbs Also See for 90207:
Table of Contents

Advertisement

After the transmission is complete and the monitor is turned off, the HS, TR and MR indicators will
always remain lit at the local modem.Setup Test
Verify that cable connections are secure.
!
Turn on the ABP monitor. It will display "9999." When the monitor is being read or initialized, the
digits will change to indicate that communication is taking place between the monitor and the
analysis system. When communication is complete, the digits will stop changing.
Office Check Mode
The monitor automatically enters an office check mode for the first five measurements immediately
following initialization. This allows you to verify the performance of the monitor on an individual
patient without the need for
re-initialization to reset the display features.
While in the office check mode, the monitor will operate as follows:
Display the cuff pressure on each bleed step
Display systole, diastole, and heart rate at the end of the measurement
Bleed one step below the diastolic value as determined by the monitor
Terminating Office Check Mode
For the 90207 (versions earlier than 2.14), press the START/STOP key twice to cancel each of the
remaining readings. The office check mode is terminated when the sum of the canceled and
successful measurements equals five.
For the 90217 and 90207 (version 2.14 and later), press the START/STOP key twice to cancel a
single blood pressure reading. The office check mode is terminated when a blood pressure reading
is cancelled.
Any event that prevents a successful blood pressure measurement (other than a manual cancel) is
not counted as one of the five office check mode readings.
Reinstating Office Check Mode
The office check mode can be reinstated in the 90217, and in versions 2.14 and later of the 90207.
To reinstate the office check mode without initialization of the monitor follow the steps below.
1.
Turn the power on to the monitor.
2.
When the version is displayed on the LCD press and hold the START/STOP key.
3.
Release the START/STOP key when EC03 is displayed on the LCD.
An EC13 will be logged to indicate the time at which the office check mode was reinstated. The
office check mode will be enabled for five additional successful measurements.
Operation
1-7

Advertisement

Table of Contents
loading

This manual is also suitable for:

90217