Thermocouple Input
Sensor Type
Type K
Type J
Type T
Type E
Type R
Type S
The supplied thermocouple probes are useable in the range of -20 to +250°C (-4 to +482°F)
only. For measuring outside of this range, please obtain thermocouples rated accordingly.
Accuracy specified for meter only. Input probes add additional measurement error.
Generally RTD probes introduce a smaller error than thermocouples.
The above specifications are tested under an environmental RF Field Strength lower than 3
V/M and a frequency lower than 30 MHz
All rights reserved including the right of reproduction in whole or in part in any form
(types K, J, T, E, R, and S)
Resolution
0.1°C
1°C
0.1°F
1°F
0.1°C
1°C
0.1°F
1°F
0.1°C
0.1°F
0.1°C
0.1°F
1°F
1°C
1°F
1°C
1°F
Copyright © 2013 FLIR Systems, Inc.
www.extech.com
Range
-50.1 to -100.0°C
-50.0 to 999.9°C
1000 to 1300°C
-58.1 to -148.0°F
-58.0 to 999.9°F
1000 to 2372°F
-50.1 to -100.0°C
-50.0 to 999.9°C
1000 to 1150°C
-58.1 to -148.0°F
-58.0 to 999.9°F
1000 to 2102°F
-50.1 to -100.0°C
-50.0 to 400.0°C
-58.1 to -148.0°F
-58.0 to 752.0°F
-50.1 to -100.0°C
-50.0 to 900.0°C
-58.1 to -148.0°F
-58.0 to 999.9°F
1000 to 1652°F
0 to 600°C
601 to 1700°C
32 to 1112°F
1113 to 3092°F
0 to 600°C
601 to 1500°C
32 to 1112°F
1113 to 2732°F
ISO‐9001 Certified
12
Accuracy
±(0.4 % + 1°C)
±(0.4 % + 0.5°C)
±(0.4 % + 1°C)
±(0.4 % + 1.8°F)
±(0.4 % + 1°F)
±(0.4 % + 2°F)
±( 0 . 4 % + 1 °C)
±(0.4 % + 0.5°C)
±(0.4 % + 1°C)
±(0.4 % + 1.8°F)
±( 0 . 4 % + 1 °F)
±(0.4 % + 2°F)
±(0.4 % + 1°C)
±(0.4 % + 0.5°C)
±(0.4 % + 1.8°F)
±( 0 . 4 % + 1 °F)
±(0.4 % + 1°C)
±(0.4 % + 0.5°C)
±(0.4 % + 1.8°F)
±( 0 . 4 % + 1 °F)
±(0.4 % + 2°F)
±(0.5 % + 3°C)
±(0.5 % + 5°F)
±(0.5 % + 3°C)
±(0.5 % + 5°F)
SDL200-EN V2.3 07/13