Specifications
General Specifications
Display
Status indicators
Measurement Channels
Sensor types
Measurement Units
Offset Adjustment
Input Linearity
Datalogger Sampling Rate
Memory Card
Temperature Compensation
Display update rate
Data Output
Operating Temperature
Operating Humidity
Auto Power OFF
Power Supply
Power Consumption
Weight
Dimensions
Electrical Specifications (ambient temperature 23°C ± 5°C)
Pt 100 ohm RTD Input (0.00385 alpha coefficient meeting DIN IEC 751)
Resolution
Backlit LCD; LCD size: 2 x 1.5"(52 x 38mm)
Over-range (----) and low battery
T1, T2, T3, T4, and T1-T2
Thermocouple types: K, J, T, E, R, and S
RTD type: platinum 100 ohm (Complies with 0.00385 alpha
coefficient meeting DIN IEC 751)
°C / °F
Available for one (1) thermocouple and two (2) RTD inputs
Sensor curves are stored in meter's CPU
AUTO LOGGING: 1, 2, 5, 10, 30, 60, 120, 300, 600, 1800,
3600 seconds. Note that a one (1) second sampling rate
can cause some data loss on slower computers
MANUAL LOGGING: Set the sampling rate to '0'
SD memory card; 1G to 16GB size
Auto temperature compensation for all thermocouple types
Approx. 1 second.
RS-232 / USB PC computer interface
32 to 122°F (0 to 50°C)
85% R.H. max.
After 10 minutes of inactivity (can be disabled)
Six (6) 1.5 VDC batteries (optional 9V AC adaptor)
Normal operation (backlight & datalogger OFF): approx. 9.7mA dc
With backlight OFF and datalogger ON: approx. 30mA dc
With backlight ON and datalogger OFF: approx. 17mA dc
0.62 lbs (280g) meter only
Main instrument: 7.1 x 2.9 x 1.9"(182 x 73 x 47.5mm)
0.1°C
-199.9 to 850.0°C
0.1°F
-327.0 to 999.9°F
1°F
Range
1000 to 1562°F
11
Accuracy
± (0.4 % + 1°C)
± (0.4 % + 1.8°F)
± (0.4 % + 2°F)
SDL200-EN V2.3 07/13
Need help?
Do you have a question about the SDL200 and is the answer not in the manual?
Questions and answers