Agilent Technologies Agilent 35900E User Manual page 18

Dual channel interface
Hide thumbs Also See for Agilent 35900E:
Table of Contents

Advertisement

2
Setup
Table 1
Configuration switch settings
Switch
Function
1
Beep/Off
2
HI/LO
3
Channel A -
TEST/NORM
4
Channel A -
ADD/IGNORE
5
Channel B -
TEST/NORM
6
Channel B-
ADD/IGNORE
Switches 1 and 2 affect the beeper operation.
The remaining four switches combine to specify either a test mode or normal
operation for each channel. The test mode provides service technicians with a
diagnostic tool to show the effects of electrical noise on a known test signal.
Switches 3 and 4 combine to specify either a test mode or normal operation
mode for Signal A.
Switches 5 and 6 combine to specify either a test mode or normal operation
mode for Signal B.
18
Position
Effect
Up
Causes an audible beep when Start/Stop is pressed.
Down
Silent at Start/Stop.
Up
Beep, if any, is loud.
Down
Beep, if any, is soft.
Up
Selects stored test chromatogram. Analog input signal ignored.
Down
Selects Analog input for Normal Operation.
Up
Add signal input to the stored chromatogram. Switch 3 must be in Up
position.
Down
Do not add (ignore) signal input to stored chromatogram for Channel A.
Up
Selects stored test chromatogram. Analog input signal ignored.
Down
Selects Analog input for Normal Operation.
Up
Add Channel B signal input to the stored chromatogram. Switch 5 must be
in Up position.
Down
Do not add (ignore) signal input to stored chromatogram for Channel B.