9.4
Replacing PWA,
Procedure, continued
176
ST 3000 Release 300 and SFC Model STS103 User's Manual
Continued
Table 55
Replacing PWA, continued
Step
4
If your transmitter...
has Local Smart Meter
Option
does not have Local Smart
Meter Option
5
Note orientation of mounting bracket on PWA (side without cable
connectors). Unplug meter cable from connector on PWA. Remove
screw retainers from other side of mounting screws so you can
remove screws and mounting bracket from PWA. Set PWA aside.
Mounting Screws
Meter Cable
Orient mounting bracket as noted above with meter cable still located
6
in slot on its right-hand side, install mounting screws through bracket
and replacement PWA, and install retainers to hold screws in place.
Plug meter cable into connector J4 on PWA and be sure cable is still
under restraining clip on front of bracket. Go to Step 9.
Action
Then...
go to Step 5
go to Step 7
Meter cable is installed
between PWA and
mounting bracket.
J4 Meter
Connector
Mounting
Bracket
Restraining
Clip
PWA Connector Side
REV
X
XXXXXXXX-XXX
ASSY
XXXXXX
R
W
Continued on next page