Display Messages and Troubleshooting (continued)
The test strip was bent or
removed during testing.
Remember to slide and release
turn the meter off and allow the
test strip to release. Slide and
release
again and a new test strip
advance.
Chapter 5: Maintenance and Troubleshooting
, at the end of the test, to
to turn on the meter
•The test strip is stuck in the test
•The meter was turned off and on
76
ACCU-CHEK Customer Care Service Center 1-800-858-8072
strip slot or the test drum does
not rotate. Pull the test strip out
of the meter and discard it. Then,
take the test drum out and
reinsert it. Remember the 90-day
use-by counter in the meter is no
longer correct. The drum symbol
flashes until you install a new
test drum.
again, but the test strip is still in
the meter. Hold the meter straight
up and down and let the test
strip drop out, or take it out
manually, even if it takes a small
amount of force.