LCD Brightness: Touch any section of the horizontal bar to select a new brightness setting. Selecting
low numbers along the bar will result in a darker screen while selecting high numbers makes the
screen appear brighter. Brightness levels range from 1 to 15.
Tone Volume: Touch any section of the horizontal bar to select a new tone volume setting. Selecting
low numbers along the bar will result in lower volume while selecting high numbers makes the
louder. Tone volume settings range from 0 to 10.
treadmill settings
Figure 7.4. The Treadmill Settings screen.
• Zero Balance: This feature is used to eliminate treadbelt "creep" (slow movement of the treadbelt
while the speed is set to zero.) Select negative values to zero out forward treadbelt creep and pos-
itive values to zero out reverse belt creep.
Note: Belt Movement While at Zero Speed
Inherent "belt creep" is common with Pulse Width Modulation Motor Control. Allow the treadmill
sufficient time to "warm up" prior to making any adjustments for belt creep.
• Incline Range: The RTM 600 comes with a choice of 0 to 15 or –3 to 12 degrees of incline. Touch
the displayed value to show both options and then touch the desired setting to select.
• Measure Units: The RTM 600 can be configured to display either U.S. or metric measurements.
To change units, touch the displayed units and then touch the desired setting to select.
• Access Protection for Custom Profiles: Access protection requires entry of an I.D. code (781) to access
custom profiles. Touch the displayed value to show both options, then touch the desired setting to
select.