Performance RHM 160
Max Flow Rate Qmax = 30,000 kg/min (66,139 lb/min) and Qnom (*) = 23,000 kg/min (50,706 lb/min)
Standard Models
Flow Rate
kg/min
30000
15000
7500
2000
750
Goldline Models (**) - selected sensors
Flow Rate
kg/min
12000
9000
7500
5000
3000
Low Flow Models (**) - selected sensors
Flow Rate
kg/min
30000
6000
2000
750
600
Repeatability
Better ± 0.1% of rate,
Goldline 0.05%
Density
Down to 0.5% uncertainty
Temperature
Better ± 1°C
(*) Nominal flow Qnom refers to roughly 8 - 10 m/s
(26 - 33 ft/s) velocity in measuring loops for best
performance.
(**) Selected sensors are only available in combination
with temperature options T1, TA, standard material and
pressure range.
Uncertainty
lb/min
in % of reading
66139
0.20
33069
0.20
16535
0.20
4409
0.20
1653
0.50
Uncertainty
lb/min
in % of reading
26455
0.15
19842
0.15
16535
0.15
11023
0.15
6614
0.15
Uncertainty
lb/min
in % of reading
66139
0.20
13228
0.20
4409
0.20
1653
0.50
1323
0.75
Uncertainty %
Flow Rate kg/min
Uncertainty %
Flow Rate kg/min
Uncertainty %
Flow Rate kg/min
• Uncertainty of reading (incl. zero drift) indications
refer to reference conditions H
1 - 3 bar (15 – 45 psi) and installation according to
field manual.
• Calibration at factory only up to 11,000 kg/min.
• Pressure drop indications refer to H
measuring loops type P1.
• For calibration to customer range and / or with
improved uncertainty, please consult your local
agent.
DP (bar)