Measurement & control versatile coriolis mass flowmeter with fast response times for very high pressure applications

Advertisement

Performance RHM 04
Max Flow Rate Qmax and Qnom (*) = 10 kg/min (22 lb/min)
Standard Models
Flow Rate
kg/min
10.0
7.0
4.0
0.5
0.2
Goldline Models (**) - selected sensors
Flow Rate
kg/min
10.0
7.0
4.0
1.0
0.5
Low Flow Models (**) - selected sensors
Flow Rate
kg/min
4.0
2.0
1.0
0.2
0.1
Repeatability
Better ± 0.1% of rate,
Goldline 0.05%
Temperature
Better ± 1°C
(*) Nominal flow Qnom refers to roughly 8 - 10
m/s (33 ft/s) velocity in measuring loops for best
performance.
(**) Selected sensors are only available in combination
with temperature options T1, TA, standard material and
pressure range.
• Serial/single path versions offer the same accuracy at
half the flow (Qmax serial version = 5 kg/min).
• No relevant pressure effect due to torsional oscillation
and semi-circle (non-deforming) measurement
section.
6
Uncertainty
lb/min
in % of reading
22.05
0.20
15.43
0.20
8.82
0.20
1.10
0.20
0.44
0.50
Uncertainty
lb/min
in % of reading
22.05
0.10
15.43
0.10
8.82
0.10
2.20
0.10
1.10
0.12
Uncertainty
lb/min
in % of reading
8.82
0.20
4.41
0.20
2.20
0.20
0.44
0.20
0.22
0.60
Uncertainty %
Flow Rate kg/min
Uncertainty %
Flow Rate kg/min
Uncertainty %
Flow Rate kg/min
• Uncertainty of reading (incl. zero drift) indications refer
to reference conditions H
bar (15 – 45 psi) and installation according to field
manual.
• Pressure drop indications refer to H
measuring loops type P1 and standard manifold block
connections.
• For calibration to customer range and / or with
improved uncertainty, please consult factory.
DP (bar)
DP (bar)
DP (bar)
O, 18-24°C (66 - 76°F), 1 - 3
2
O, with parallel
2