Specifications - Fluke 1730/US User Manual

Energy
Hide thumbs Also See for 1730/US:
Table of Contents

Advertisement

Specifications

General Specifications
Color LCD Display ................................................ 4.3-inch active matrix TFT, 480 pixels x 272 pixels, resistive touch panel. Text and graphics in color.
Power/Charging/LED Indicator
Warranty
1730 and Power Supply ..................................... 2 years (battery not included)
Accessories ........................................................ 1 year
Calibration Cycle .................................................. 2 years
Dimensions
1730 .................................................................... 19.8 cm x 16.7 cm x 5.5 cm (7.8 in x 6.6 in x 2.2 in)
Power Supply ..................................................... 13.0 cm x 13.0 cm x 4.5 cm (5.1 in x 5.1 in x 1.8 in)
1730 with power supply attached ....................... 19.8 cm x 16.7 cm x 9 cm (7.8 in x 6.6 in x 3.5 in)
Weight
1730 .................................................................... 1.1 kg (2.5 lb)
Power Supply ..................................................... 400 g (0.9 lb)
External Protection ............................................... Holster, Kensington lock
Environmental Specifications
Operating Temperature ........................................ 10 C to 50 C (14 F to 122 F)
Storage Temperature............................................ 20 C to 60 C ( 4 F to 140 F), with battery: 20 C to 50 C ( 4 F to 122 F)
Operating Humidity .............................................. <10 °C (<50 °F) non condensing
Operating Altitude ................................................ 2,000 m (up to 4,000 m derate to 1000 V CAT II/600 V CAT III/300 V CAT IV)
Storage Altitude .................................................... 12,000 m
IP Rating ................................................................ IEC 60529:IP50, in connected condition with protection caps in place.
Vibration ................................................................ MIL 28800E, Type 3, Class III, Style B
Safety ..................................................................... IEC 61010-1: Overvoltage CAT IV, Measurement 1000 V CAT III / 600 V CAT IV, Pollution Degree 2
Electromagnetic Environment ............................. IEC 61326-1: Industrial
Electromagnetic Compatibility ............................ Applies to use in Korea only. Class A Equipment (Industrial Broadcasting & Communication Equipment)
Shop for Fluke products online at:
10 °C to 30 °C (50 °F to 86 °F) 95 %
30 °C to 40 °C (86 °F to 104 °F) 75 %
40 °C to 50 °C (104 °F to 122 °F) 45 %
[1]
This product meets requirements for industrial (Class A) electromagnetic wave equipment and the seller or user
should take notice of it. This equipment is intended for use in business environments and not to be used in homes.