NOTE: The Transducer Sensitivity (5 uV/V/mmHg or 40 uV/V/mmHg) must be set to
correlate with the monitoring equipment before simulation can begin. (See SETUP for
selection information).
The PS-2110 offers one Blood Pressure Channel and will simulate the set Blood Pressure
wave during ECG waveforms where it occurs.
There are 14 Blood Pressure settings available. Each of the six dynamic waveforms will
synchronize with the NSR rate or arrhythmia selection.
The display will resemble the following:
WAVEFORM
Arterial
120/80
Left Vent
120/0
Right Vent
25/0
Pulm Artery
25/10
CVP
15/10
PAW
10/2
Static
0 mmHg*
Static
20 mmHg
Static
40 mmHg
Static
80 mmHg
Static
100 mmHg
Static
200 mmHg
Static
250 mmHg
Static
300 mmHg
Auto Static Pressure

BLOOD PRESSURE

* Indicates Default Setting
(See Power Up Settings)