SECTION 8—TROUBLESHOOTING GUIDE
NO CHARGING SYMBOL SHOWN ON STANDBY SCREEN
SYMPTOM
With the unit turned
Power source is no good, or
Off and AC or DC
there is a loose connection.
power plugged into
the unit, no charging
Battery is outside the allowed
icon is displayed.
temperature range for charging.
Battery module is not fully
seated.
Standby screen is not being
displayed.
UNIT IS BEEPING WITH NO LIGHTS OR DISPLAY
SYMPTOM
All power sources are
Battery pack has been removed
missing.
and there is no external power
While operating without a
battery, the external power is
lost.
Invacare SOLO
2
PROBABLE CAUSE
PROBABLE CAUSE
36
SOLUTION
Try another power outlet
and check connections to
unit.
Allow unit to cool down to
less than 95° F
(35° C), or warm up to 50° F
(10° C).
Ensure the battery module is
completely inserted.
Press the On/Off button to
display the Standby screen
when using DC power.
SOLUTION
Install any battery pack
available.
Beeping will last 10-15
minutes if no power source
is provided.