Configure Dcv Menu Structure; Table - Keithley 2002 User Manual

Hide thumbs Also See for 2002:
Table of Contents

Advertisement

Front Panel Operation

Table 2-6

CONFIGURE DCV menu structure

Menu item
SPEED
NORMAL
FAST
MEDIUM
HIACCURACY
SET-SPEED-EXACTLY
SET-BY-RSLN
FILTER
AUTO
AVERAGING
ADVANCED
AVERAGING-MODE
RESOLUTION
AUTO
3.5d, 4.5d, 5.5d, 6.5d,
7.5d, 8.5d
2-10
Description
Measurement speed (integration time) menu:
Select 1 PLC (power line cycle, 16.67msec for 60Hz, 20msec for 50Hz and 400Hz).
Select 0.01 PLC.
Select 0.1 PLC.
Select 10 PLC.
Set integration time in PLC (0.01-50).
Default to setting appropriate for resolution.
Digital filter menu:
Default to filter appropriate for integration time.
Program a simple average filter (1-100 readings).
Program a simple average filter (1-100 readings) with a noise tolerance window (0-
100% of range).
Select moving average or repeating average mode.
Display resolution menu:
Default to resolution appropriate for integration time.
Select a specific resolution.

Advertisement

Table of Contents
loading

Table of Contents