The Medalist x6000 is powered by 200-240VAC Wye or
Delta or 220/380-240/415 VAC wye with neutral, 3-phase,
50/60Hz (cycles) AC mains. The PDU provides an output of
200-240VAC 50/60Hz to the Medalist x6000 subsystems.
The PDU can be reconfigured based on the AC mains input
to provide the appropriate AC voltage to the Medalist x6000
subsystems. The PDU outputs are protected with either a
10A 250VAC fuse or a 15A 250VAC thermal circuit breaker.
The PDU provides emergency stop control via a 12Vdc
control circuit. The PDU will disconnect power to all
switched subsystems via an internal contactor by:
Pressing one of the emergency stop push buttons
Opening the AC mains disconnect circuit breaker
supplying power to the Medalist x6000.
The emergency stop circuit will not power off the system
controllers or monitor.