Operation
Description:
nnnnnn - Base number for Ko1; nnnnnn (000000 to 999999) represents a number from 0 to
0.999999. Ko1 controls the cut off frequency of the offset error noise filter.
oooooo - Base number for Ko2; oooooo (000000 to 999999) represents a number from 0 to
0.999999. Ko2 controls the cut off frequency of the frequency error noise filter.
Response Format:

Enter Disgain

Command Format:
Description:
ddddd - Base number for Disgain; ddddd (00000 to 99999) controls gain common to both the
offset error and the frequency error paths of the discipline loop.
Response Format:

Enter Resync Threshold

Command format:
Description:
rrrrrr.rrr - Resync Threshold; rrrrrr.rrr (microseconds) is the limit value that determines if a
resync will occur when disciplining restarts. A resync will always be occur after power up.
Response Format: