Measurement Operation(1)

A. DC Voltage Measurement
Warning
To avoid harms to you or damages to the Meter from
eletric shock, do not attempt to measure voltages
higher than 600V AC/DC, although readings may be
obtained.
The DC Voltage ranges are:
200.0mV, 2.000V, 20.00V, 200.0V and 600V.
To measure DC voltage, connect the Meter as follows:
1. Insert the red test lead into the
and the black test lead into the COM terminal.
2. Set the rotary switch to V
3. Connect the test leads across with the object being
measured.
The measured value shows on the display.
Note
l In each range, the Meter has an input impedance of
10M . This loading effect can cause measurement
errors in high impedance circuits. If the circuit impedance
is less than or equal to 10k
(0.1 or less).
Model UT201/202: OPERATING MANUAL
(see figure 3)
red
black
( figure 3)
V terminal
.
the error is negligible