Agilent MOI for DisplayPort Cable & Connector Compliance Tests
Connector
Wire management
Cable
Impedance Profile Measurement Impedance Limits & Connector Profile Example
Impedance Profile Through Mini DisplayPort Connector
Segment
Fixture
Connector
Wire Management
Cable
100 ohm
+-10 %
100 ohm
+-10%
100 ohm
+-5 %
Differential
Maximum
Impedance Value
Tolerance
100 ohm
+- 10 %
100 ohm
+- 15 %
100 ohm
+-10 %
100 ohm
+-5 %
Transition from +-10% to +- 5% must
have a slope of 5 ohm/80ps
Comment
Fixture should have trace lengths of no
more than 50 mm (2-inches)
Transition from +-15% to +-10% and
transition from +/-10% to +/-5% shall
have a slope of 10ohm/200ps
40