e9321_90001.book Page 13 Tuesday, March 18, 2003 10:44 AM

The E-Series E9320 Power Sensors in Detail

The E-Series E9320 Power
Sensors in Detail
The E-series E9320 power sensors have two frequency
ranges. The E9325A, E9326A, and E9327A have a frequency
range of 50 MHz to 18 GHz while the 50 MHz to 6 GHz range of
the E9321A, E9322A, and E9323A covers most wireless
communication applications.
The sensors have two independent measurement paths as
shown in Figure 1.
RF IN
50ohm
Load Filter*
(300 kHz, 1.5 MHz
5 MHz low pass)
Sensor Diode
Bulkhead
* Bandwidth is sensor dependent
Figure 1 Simplified Sensor Block Diagram
Use the default normal path for continuously sampled
measurements of modulated signals and time gated
measurements. For each frequency range there is a choice of
sensors with three video (modulation) bandwidths.
E-Series E9320 Operating and Service Guide
Average Only Path
Chopper
CW Differential
AMP
Selectable Gain
Normal
Differential Amplifier*
Path
(300 kHz, 1.5 MHz, 5 MHz)
PEAK AUTO-ZERO
GAIN SELECT
Thermistor Bias
2
I
C Buffer
Gain/Mode Control
Sensor ID
2
E
PROM
Switched
Average
Gain
Only Path
Preamp
CHOP/CLOCK
CHOP/GAIN
Normal
Path
SERIAL BUS