Probe Specifications - Teledyne 9060L Operating Instructions Manual

Percent oxygen analyzer
Table of Contents

Advertisement

Appendix
A.2 Probe Specifications
Application: Combustion flue below 700°C (1290°F),
Temperature Range: 0-900°C
Length: 250-1000 mm 500-1160 mm 457-1219 mm
Process Connection: 1 1/2" BSP 3/4" BSP 1" BSP or equivalent
Electrical Connection: Weatherproof plug-in connector or
Heater: Yes
Flue Gas Thermocouple: K
Response Time: Typically < 4 secs.
Head Temperature: 125°C (250°F)
Reference Gas: Ambient air 50 to 500 cc/min (6 to 60
Ref Air Connection: 1/4" NPT integral air line in probe cable.
Calibration Gas Connection: 1/8" NPT female 1/8" NPT female 1/8"
Weight : 0.6 kg (1.32 lbs.)
Teledyne Analytical Instruments
above 700°C (1290°F) 900°C (1650°F) — no
contaminants.
With contaminants see notes 1 &2
(32-1650°F) (1470-2550°F) (1470-2190°F)
(10" to 40")
NPT
optional screw terminals. Plug connector
supplied with the cable. Ex d heads have
screw terminals.
Cable: Order a specific length with the analyzer
except for hazardous installations where
the cable is supplied by the customer.
scfm). Pump supplied with analyzer
Barbed fitting to 3/16" ID PVC tube.
Filter: Removable sintered stainless steel
particulate filter, 15 micron.
NPT female
Model 9060L
800-1400°C
800-1200°C
(20" to 46")
(18" to 48")
44

Advertisement

Table of Contents
loading

Table of Contents