Percent Oxygen Analyzer

Appendix

A.1 Specifications
Ranges: Two user selectable % ranges can be set
Signal Output: Voltage: 0–10 VDC, negative ground
Range ID: 0-10 VDC
Display: Light emitting diode.
Alarms: One high alarm relay, adjustable; one low
System Operating Temp: 32-122°F (0-50 °C)
Accuracy: ±2% of full scale at constant temperature
Response Time: 90% in less than 20 seconds at 77°F
System Power Requirements: AC (100 to 240 VAC, 50/60 Hz), or
Power Consumption: 70 Watts
Teledyne Analytical Instruments
between 1% and 25 % O
cal range. (Optional 0-100% range is
available. Consult factory.)
Standard ranges configured as 0-3% and 0-
10%.
Sensor: E-2 Oxygen Sensor (standard)
C-5, C-5F Oxygen Sensor (optional)
Current: 4-20 mA, negative ground
alarm relay, adjustable; one sensor failure
relay. (All are failsafe.)
and pressure
±5% of full scale through operating
temperature range once temperature
equilibrium is reached and at constant
pressure
(25 °C)
DC (10-36 VDC, 15 W); user specified
Maintenance
plus fixed 0-25%
2
35