Specifications, Agency Approvals, Warranty; Specifications - Teledyne 6200A Manual

Trs analyzer
Table of Contents

Advertisement

TAI Model 6200A SO
3 SPECIFICATIONS, AGENCY APPROVALS,
WARRANTY

3.1 Specifications

Ranges
Noise at Zero
Noise at Span
Lower Detectable Limit
3
Zero Drift
3
Zero Drift
3
Span Drift
Lag Time
Rise Time
Fall Time
Sample Flow Rate
Linearity
Precision
Temperature Range
Temp Coefficient
Humidity
Voltage Coefficient
Dimensions HxWxD
Weight, Analyzer
Power, Analyzer
Power, Analyzer
Environmental Conditions
Recorder Output
Status Option
Measurement Units
1. As defined by USEPA.
2. Defined as twice the zero noise level.
3. At constant temperature and voltage.
4. Electrical ratings for CE Mark compliance
5. Bi-polar.
Analyzer Operator Manual, 02164, Rev. G
2
In 1ppb increments from 50ppb to 20,000ppb, dual ranges or
autoranging
1
0.2 ppb RMS
1
<0.5% of reading (above 50 ppb)
2
0.4 ppb RMS
<0.5 ppb/24 hours
1 ppb/7 days
<0.5% FS/24hrs
20 sec
95% in <120 sec
95% in <100 sec
700cc/min. ±10%
1% of full scale
0.5% of reading
5-40
< 0.1% per
0 - 95% RH, non-condensing
< 0.05% per V
7"x17" x 23.6" (18cm x 43cm x 61cm)
45 lbs (20.5 kg) w/internal pump
110v/60Hz, 220v/50Hz, 240v/50Hz, 250 watts
4
230v/50Hz, 2.5A
Installation Category (Overvoltage Category) II
Pollution Degree 2
5
0-100 mV, 0-1, 5, 10v ; resolution of 1 part in 1024 of selected
voltage or current range
12 Status Outputs from opto-isolator
ppb, ppm, ug/m
1
o
C
o
C
3
3
, mg/m
3-1

Advertisement

Table of Contents
loading

Table of Contents