Operation
V1.0
-----------------------
Entering the MODEL screen causes the above screen to be
displayed for 3 seconds. There are no modifiable parameters.

4.3.6 Self-Test Routine

-----------------------
Begin Self-Test?
ENTER-YES ESCAPE-NO
-----------------------
Pressing ENTER causes the analyzer to run the same diagnostic
that runs on start-up. See section 5.4 for results display.

4.3.7 Filter Routine

-----------------------
Digital Filter
->Index: 5
-----------------------
Pressing ENTER causes the number after 'Index:' to start
flashing. Then pressing UP or DOWN selects a number ranging from 1
to 10. Lower filter values cause the analyzer to converge on new inputs
more quickly, but also allow more noise in the sample.

4.3.8 Analog-Out Adjust Routine

-----------------------
4-20mA Offset: 0
Teledyne Analytical Instruments
3000ZA-3X
34