Externally Applied Voltages

Caution - Front Panel Sockets
To avoid risk of damage to the instrument:
Do not apply external voltages to the MAIN
OUT or AUX OUT sockets.
Do not apply external voltages exceeding ±10
V to the TRIG IN socket.
Caution - Rear Panel Sockets
To avoid risk of damage to the instrument:
Do not apply external voltages to the
SYNC OUT or TRIG/SWEEP OUT sockets.
Do not apply external voltages exceeding
±10 V to the VCA IN socket.
Do not apply external voltages exceeding
+7.5 V or -2.5 V to the CLOCK IN/OUT socket.