Ph Values At Various Temperatures; Ec/Tds Calibration - Hanna Instruments HI 9813-5 Instruction Manual

Portable ph/ec/tds/°c meters
Hide thumbs Also See for HI 9813-5:
Table of Contents

Advertisement

Note: The probe should be submerged approximately 4 cm (1½") into the
solution. The ChecktempC thermometer has to be located close to
the probe.
Note: If turning the knob the needed value can not be reached,
clean the probe (see the "Probe Maintenance" section).
If also after the probe cleaning the value can not be reached,
replace the probe.

pH VALUES AT VARIOUS TEMPERATURES

For temperature compensation during calibration, please refer to the
following chart.
TEMP
°C
°F
4.01
6.86
0
32
4.01
6.98
5
41
4.00
6.95
10
50
4.00
6.92
15
59
4.00
6.90
20
68
4.00
6.88
25
77
4.01
6.86
30
86
4.02
6.85
35
95
4.03
6.84
40
104
4.04
6.84
45
113
4.05
6.83
50
122
4.06
6.83
55
131
4.07
6.84
60
140
4.09
6.84
65
149
4.11
6.85
70
158
4.12
6.85
For instance, if the buffer temperature is 25°C, the display should
show pH 4.0 or 7.0 or 10.0.
If the buffer temperature is 10°C, the display should show pH 4.0 or
7.0 or 10.1.
10
pH VALUES
7.01
9.18
10.01
7.13
9.46
10.32
7.10
9.39
10.24
7.07
9.33
10.18
7.04
9.27
10.12
7.03
9.22
10.06
7.01
9.18
10.01
7.00
9.14
9.96
6.99
9.10
9.92
6.98
9.07
9.88
6.98
9.04
9.85
6.98
9.01
9.82
6.98
8.99
9.79
6.98
8.97
9.77
6.99
8.95
9.76
6.99
8.93
9.75

EC/TDS CALIBRATION

Accessories needed:
• Use HI 70031 (1413 µS/cm = 1.413 mS/cm) EC calibration
solution or HI 70442 (1500 ppm) TDS calibration solution.
Note: The conversion between EC and TDS is made by a built-in
circuit, hence it is requested to calibrate the meter only in EC
or TDS range. The other range is thus automatically cali-
brated.
PROCEDURE
• Pour approximately 4 cm (1½") of a conductivity calibration
solution (e.g. HI 7031) into a beaker. If possible, use plastic
beaker to minimize any EMC interference.
• Immerse the probe in the solution.
• Wait for a couple of minutes for thermal equilibrium to be
reached.
• Tap the probe on the bottom, then shake it lightly while rotating
to make sure no air bubbles remain trapped inside the probe.
• Press the "mS/cm" (or "ppm") key.
• Turn the EC/TDS calibration knob until
the display shows the EC or TDS reading
at 25°C.