Hanna Instruments HI 91610 Instruction Manual

Hanna Instruments HI 91610 Instruction Manual

Printing and logging thermo-hygrometers
Hide thumbs Also See for HI 91610:

Advertisement

Quick Links

Instruction Manual
HI 9161 • HI 91610
Printing and Logging
Thermo-Hygrometers
w w w . h a n n a i n s t . c o m

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HI 91610 and is the answer not in the manual?

Questions and answers

Summary of Contents for Hanna Instruments HI 91610

  • Page 1 Instruction Manual HI 9161 • HI 91610 Printing and Logging Thermo-Hygrometers w w w . h a n n a i n s t . c o m...
  • Page 2: Table Of Contents

    Dear Customer, Thank you for choosing a HANNA instruments ® product. All Hanna Instruments meters are warranted for two years Please read this instruction manual carefully before using the instru- against defects in workmanship and materials when used for their ment.
  • Page 3: General Description

    GENERAL DESCRIPTION FUNCTIONAL DESCRIPTION HI 9161 HI 9161 and HI 91610 are advanced portable, printing, thermo- hygrometers. These unique, lightweight meters give instant printouts of sample number, date and time of recording, printing interval, relative humidity and temperature at user-selectable intervals. Both RH and temperature are measured with 0.1 resolution.
  • Page 4: Functional Description Hi 91610

    FUNCTIONAL DESCRIPTION FUNCTIONAL DESCRIPTION HI 91610 DISPLAY AND RH PROBE HI 91610C MICROPROCESSOR LOGGING THERMOHYGROMETER RANGE 5.0 . . . 95.0 °C -20.0 . . +60.0 PRINT PAPER DOWN RANGE TIME 1. Power supply socket 1. Probe shielded cable 2. RH probe socket 2.
  • Page 5: Specifications

    INITIAL PREPARATION SPECIFICATIONS Each meter is supplied complete with four 1.5V AA batteries. Unscrew HI 9161C HI 9161F and remove the back cover, unwrap the batteries and install them HI 91610C HI 91610F while paying attention to their polarity (see "Battery Replacement" Range 5.0 to 95.0% RH section for details).
  • Page 6: Operational Guide

    SETTING DATE, TIME & PRINTING INTERVAL OPERATIONAL GUIDE WITH HI 9161 When the instrument is turned on, the dis- RELATIVE HUMIDITY MEASUREMENTS play shows the TIME. TIME INTV Operations with the HI 9161 and HI 91610 are very simple. • For accurate measurements, the RH sensor should be exposed to a Press the INTV and RANGE keys simultaneously.
  • Page 7 The time, date and printing interval are now stored in the memory Press the INTV and RANGE keys simultaneously. The display will show even when the meter or display are switched off. the time/printing interval setting. The printing interval on the secondary LCD will blink.
  • Page 8 Press the ALT and TIME keys simultaneously and the display will The time, date and printing interval are now stored in the memory show the previously memorized time with the printing interval even when the meter or display are switched off. blinking on the secondary LCD.
  • Page 9: Printing/Recording With Hi 9161

    VIEWING RH%, TEMPERATURE, TIME, DATE & PRINT- PRINTING/RECORDING WITH HI 9161 ING INTERVAL WITH HI 91610 To view the time press the TIME key. This also displays the printing/ To print the measured values press PRINT. logging interval. Each printout provides the following informa- tion: PRINT TIME...
  • Page 10: Printing/Logging With Hi 91610

    PRINTING/LOGGING WITH HI 91610 Press the RANGE key to read the temperature on the primary display. RANGE To print the values shown on the display, press PRINT. If no key is pressed, after approximately 5 minutes, the meter goes in PRINT standby mode to save the battery life.
  • Page 11 Each printout provides the following information: To reactivate the display press the TIME key. TIME Notes: 1 4 - 0 1 - 9 6 • Once in the logging mode, the interval cannot be changed. Exit - - 0 1 - - 0 0 0 1 M the logging mode (press ALT and LOG together) and then set the 0 0 1 5...
  • Page 12 TO STOP LOGGING Press UP and the RH% value will be displayed on the primary LCD. Press the ALT and LOG keys simultaneously. This will also generate an exit status printout. 01-07-95 LOG EXIT Log number ==02== 0005M Date/Time <>0010 *7.01 Total sample number Continue pressing UP will display one by one all the memorized data of...
  • Page 13: Calibration

    All HANNA thermo-hygrometers have been precalibrated at the ® factory. HANNA instruments uses state-of-the-art thermally-isolated • Pour approximately 12 cc of distilled wa- humidity chambers for this purpose.
  • Page 14: Data Transfer To Pc (Hi 91610 Only)

    • Adjust the low humidity trimmer to read 0.0% (readings between DATA TRANSFER TO PC 0.0 and 1.0% RH are acceptable). (HI 91610 only) Set the HI 91610 to TIME mode and place it on a HI 9200 infrared transmitter (ensuring that the infrared LEDs of meter and interface TIME match).
  • Page 15: Self-Diagnostic Functions

    BATTERY REPLACEMENT SELF-DIAGNOSTIC FUNCTIONS HI 9161 and HI 91610 are factory programmed to automatically When the batteries become weak, diagnose a fault and inform the user by displaying an error code on the "LO BAT" indication is displayed the LCD. to warn the user.
  • Page 16: Printer Maintenance

    Allow about 5 cm (2") of paper to exit from the printer and then PRINTER MAINTENANCE replace the cover. CHANGING THE INK CARTRIDGE When printouts become faint, it might be necessary to change the ink cartridge. Contact your dealer or the nearest HANNA Office for technical service.
  • Page 17: Accessories

    HANNA INSTRUMENTS TEMPERATURE PROBES ACCESSORIES RH PROBES HANNA Relative Humidity probes use a high-tech TFPC (Thin-Film Polymer Capacitance) humidity sensor, that provides rapid response and high accuracy. The probes are precalibrated at the factory. Several different versions are available for all application needs.
  • Page 18: Ce Declaration Of Conformity

    Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance. Unplug the instruments from power supply before replacing the fuse or making any electrical connections. Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
  • Page 19 SALES AND TECHNICAL SERVICE CONTACTS Australia: Tel. (03) 9769.0666 • Fax (03) 9769.0699 China: Tel. (10) 88570068 • Fax (10) 88570060 Egypt: Tel. & Fax (02) 2758.683 Germany: Tel. (07851) 9129-0 • Fax (07851) 9129-99 Greece: Tel. (210) 823.5192 • Fax (210) 884.0210 Indonesia: Tel.

This manual is also suitable for:

Hi 9161

Table of Contents