Hanna Instruments HI 91610 Instruction Manual
Hanna Instruments HI 91610 Instruction Manual

Hanna Instruments HI 91610 Instruction Manual

Portable microprocessor printing and logging thermohygrometers

Advertisement

Quick Links

Instruction Manual
HI 9161 - HI 9261
HI 91610
Portable
Microprocessor
Printing and Logging
Thermohygrometers
These Instruments are in
http://www.hannainst.com
Compliance with the CE Directives

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HI 91610 and is the answer not in the manual?

Questions and answers

Summary of Contents for Hanna Instruments HI 91610

  • Page 1 Instruction Manual HI 9161 - HI 9261 HI 91610 Portable Microprocessor Printing and Logging Thermohygrometers These Instruments are in http://www.hannainst.com Compliance with the CE Directives...
  • Page 2: Table Of Contents

    GENERAL DESCRIPTION SPECIFICATIONS................9 INITIAL PREPARATION ............... 10 OPERATIONAL GUIDE ..............11 HI 9161, HI 9261 and HI 91610 are portable, microprocessor-based PRINTING/RECORDING WITH HI 9161 & HI 9261 ...... 18 printing/logging thermohygrometers. PRINTING/LOGGING WITH HI 91610 .......... 20 The housing is made of rugged, lightweight material, making them CALIBRATION ................
  • Page 3: Functional Description Hi 9161

    HI 9200 transmitter. The internal software allocates memory space to store up to 8000 measurements. HI 91610 will transfer the data in seconds through the infrared lights with no need for a cable between the transmitter and the meter.
  • Page 4: Functional Description Hi 9261

    FUNCTIONAL DESCRIPTION HI 91610 FUNCTIONAL DESCRIPTION HI 91610 FUNCTIONAL DESCRIPTION HI 91610 FUNCTIONAL DESCRIPTION HI 91610 FUNCTIONAL DESCRIPTION HI 91610 FUNCTIONAL DESCRIPTION HI 9261 FUNCTIONAL DESCRIPTION HI 9261 FUNCTIONAL DESCRIPTION HI 9261 FUNCTIONAL DESCRIPTION HI 9261 FUNCTIONAL DESCRIPTION HI 9261...
  • Page 5: Specifications

    F U N C T I O N A L D E S C R I P T I O N F U N C T I O N A L D E S C R I P T I O N F U N C T I O N A L D E S C R I P T I O N F U N C T I O N A L D E S C R I P T I O N F U N C T I O N A L D E S C R I P T I O N...
  • Page 6: Initial Preparation

    The HI 762L/2 temperature probe supplied with the HI 9161, To maximize battery life, the display is automatically switched off after HI 9261 and HI 91610 is a liquid/general purpose probe with 2 m 5 minutes of non-use. However, the meter will continue to monitor (6.6') cable.
  • Page 7 If neither a temperature probe nor a calibra- When the correct year is selected, press the RANGE key once and the tion test plug is connected, the meter will month will start blinking. display on the primary display and print "----"...
  • Page 8 RANGE TIME HI 91610 Any interval can be selected from 1, 2, 5, 10, 15, 30, 60, 120 or 180 minutes by using the UP and DOWN keys. Press the ALT and the TIME keys simultaneously. The display will show the previously memorized date with the year blinking on the secondary display (as the last two digits of the year, e.g.
  • Page 9 DATE DOWN TO VIEW RH%, TEMPERATURE, TIME, DATE & PRINTING INTERVAL HI 91610 To view the time press the TIME key. This also displays the printing/ HI 9161 & HI 9261 logging interval. TIME...
  • Page 10 To view RH% on the primary display press RANGE. meter will print the measurements taken in that moment, and will print with the indi- cated printing interval thereafter until the INTV ON/OFF key is pressed. The printout provides the following information: RANGE a –...
  • Page 11 After approximately 5 minutes the display will switch itself off but the logging function remains active. To reactivate the display press the TIME key. PRINTING/LOGGING PRINTING/LOGGING PRINTING/LOGGING WITH WITH WITH WITH HI HI HI HI HI 91610 91610 91610 91610 PRINTING/LOGGING PRINTING/LOGGING WITH 91610...
  • Page 12 Press the ALT and LOG key simultaneously to enter the logging mode. Notes: The current log number and the remaining page numbers will • It is recommended to use an external power supply during appear for a few seconds on the display to indicate the correct logging with printing mode, especially when many printouts are operational mode.
  • Page 13 While pressing the ALT key, press the DOWN key until the log number to scan appears on the secondary display. The primary display will To exit from the recall mode press the LOG key. show the number of samples in that particular log. Note: this mode will not alter data already present in the memory.
  • Page 14: Calibration

    All Hanna thermohygrometers have been precalibrated at the factory. • Seal the chamber well when not in use, Hanna Instruments uses state-of-the-art thermally-isolated humidity as the LiCl solution is extremely hygro- Na Cl chambers for this purpose.
  • Page 15: Data Transfer To Pc (Hi 91610)

    • Leaving the probe in the "RH 75.4%" chamber, adjust the low HI 9261) or the RANGE key (for humidity trimmer until the display shows 75.4%. HI 91610) to switch the instrument on. ON/OFF • Press the RANGE key to display the RH% reading on the primary LCD.
  • Page 16: Self-Diagnostic Functions

    HI 9161, HI 9261 and HI 91610 are factory programmed to automatically diagnose a fault and inform the user by displaying an HI 9161, HI 9261 and HI 91610 use plain 38 mm wide paper rolls. error code on the LCD.
  • Page 17: Battery Replacement

    It is recommended to use the Hanna HI 710005 and HI 710006 voltage adapters that use the proper polarity configuration. HI 9161, HI 9261 and HI 91610 can also be used with other adapters supplying a 12VDC output. In this case, remember to check...
  • Page 18: Accessories

    R.H. PROBES HI 762PR Penetration probe with 1 m (3.3') cable and red Hanna Instruments Relative Humidity probes use a high-tech Thin- handle Film Polymer Capacitance humidity sensor. This sensor provides rapid HI 762PR/10 Penetration probe with 10 m (33') cable and red response and high accuracy.
  • Page 19: Warranty

    Calibration key, +158.0°F ±0.8°F is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Cus- tomer Service department and then send it with shipment costs prepaid.
  • Page 20: Ce Declaration Of Conformity

    • DISSOLVED OXYGEN METERS • HYGROMETERS DECLARATION OF CONFORMITY • ION SPECIFIC METERS (Colorimeters) • MAGNETIC STIRRERS • Na/NaCl METERS Hanna Instruments Srl V.le delle industrie 12 • pH/ORP/Na ELECTRODES 35010 Ronchi di Villafranca (PD) ITALY • pH METERS herewith certify that the printing/logging thermohygrometers...
  • Page 21 HANNA LITERATURE HANNA LITERATURE HANNA LITERATURE HANNA LITERATURE HANNA LITERATURE LAB RECORDING WATER ANALYSIS ENVIROCARE GENERAL CATALOG These and many others catalogs, handbooks and leaflets are avail- able from Hanna. To receive your free copy, contact your dealer or the nearest Hanna Customer Service Center.