Instruction Manual HI 9033 - HI 9034 HI 9039 - HI 9130 HI 933100 - HI 9635 Portable Waterproof Multi-Range Conductivity/TDS/ Fertilizer Meters These Instruments are in M a n u f a c t u r e r s...
GENERAL DESCRIPTION ..............3 FUNCTIONAL DESCRIPTION OF HI 76302W CONDUCTIVITY PROBE ..4 FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 9033 ..5 FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 9034 ..6 FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 9039 ..7 FUNCTIONAL DESCRIPTION &...
cally compensated for by the ATC circuitry of the meters. The stainless steel rings of the probe have a large surface for better response and easy cleaning. The detachable probe can be changed without returning the unit to the factory. The probe cable is screened and 1 m (3.3') long.
FUNCTIONAL DESCRIPTION & FUNCTIONAL DESCRIPTION & FUNCTIONAL DESCRIPTION & FUNCTIONAL DESCRIPTION & FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 9033 SPECIFICATIONS OF HI 9033 SPECIFICATIONS OF HI 9033 SPECIFICATIONS OF HI 9033 SPECIFICATIONS OF HI 9033 1) Battery Compartment 2) Probe Socket 3) LCD Display 4) "mS"...
FUNCTIONAL DESCRIPTION & FUNCTIONAL DESCRIPTION & FUNCTIONAL DESCRIPTION & FUNCTIONAL DESCRIPTION & FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 9039 SPECIFICATIONS OF HI 9039 SPECIFICATIONS OF HI 9039 SPECIFICATIONS OF HI 9039 SPECIFICATIONS OF HI 9039 1) Battery Compartment 2) Probe Socket 3) LCD Display 4) ON/OFF Key 5) RANGE, to select...
FUNCTIONAL DESCRIPTION & FUNCTIONAL DESCRIPTION & FUNCTIONAL DESCRIPTION & FUNCTIONAL DESCRIPTION & FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 9130 SPECIFICATIONS OF HI 9130 SPECIFICATIONS OF HI 9130 SPECIFICATIONS OF HI 9130 SPECIFICATIONS OF HI 9130 1) Battery Compartment 2) Probe Socket 3) "mS"...
FUNCTIONAL DESCRIPTION & FUNCTIONAL DESCRIPTION & FUNCTIONAL DESCRIPTION & FUNCTIONAL DESCRIPTION & FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 933100 SPECIFICATIONS OF HI 933100 SPECIFICATIONS OF HI 933100 SPECIFICATIONS OF HI 933100 SPECIFICATIONS OF HI 933100 1) Battery Compartment 2) Probe Socket 3) Liquid Crystal Display 4) ON/OFF Key 5) CAL, to enter or exit cali-...
FUNCTIONAL DESCRIPTION & FUNCTIONAL DESCRIPTION & FUNCTIONAL DESCRIPTION & FUNCTIONAL DESCRIPTION & FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI 9635 SPECIFICATIONS OF HI 9635 SPECIFICATIONS OF HI 9635 SPECIFICATIONS OF HI 9635 SPECIFICATIONS OF HI 9635 1) Battery Compartment 2) Probe Socket 3) Liquid Crystal Display 4) ON/OFF Key 5) CAL, to enter or exit calibration mode...
Page 11
RANGE Conductivity 0.0 to 150.0 µS/cm 150 to 1500 µS/cm 1.50 to 15.00 mS/cm 15.0 to 199.9 mS/cm 0.0 to 150.0 mg/L (ppm) 150 to 1500 mg/L (ppm) 1.50 to 15.00 g/L (ppt) 15.0 to 100.0 g/L (ppt) Temperature 0.0 to 60.0°C Resolution Cond.
OPERATIONAL GUIDE OPERATIONAL GUIDE OPERATIONAL GUIDE OPERATIONAL GUIDE OPERATIONAL GUIDE INITIAL PREPARATION Each meter is supplied complete with battery(ies). Remove the back cover, unwrap the battery(ies) and install it (them) while paying attention to polarity its (their) (see page 31). Connect the probe to the meter securely by aligning the pins with the socket, pushing the plug in and tightening the threaded ring.
Page 13
HI 933100 and HI 9635 are auto-ranging conductivity meters, and the reading automatically switches from one resolution to the next. With HI 9033, HI 9034 and HI 9130, if the LCD displays a "1" on the far left hand side, the meter is out of range and the solution measured exceeds the selected range.
Page 14
Select the temperature (°C) mode by pressing FACTOR once. The secondary LCD figure will display a default FACTOR temperature of 25°C or the last recorded tem- perature reading with a blinking "°C" symbol. Change the reading by pressing the UP or °C DOWN keys.
A complete list of the Hanna calibration solutions is given in the Accessories section (page 32). PROCEDURE FOR HI 9033 AND HI 9130 • Fill a beaker with 8 cm (3¼") of conduc- tivity calibration solution (if possible fill two beakers and use one as rinse and the other for calibration).
Page 16
minutes if there is a temperature difference of 5°C (9°F) or more for the ATC circuitry to compensate completely. • When the reading stabilizes, turn the calibration trimmer until the display reads the proper conductivity value @25°C. For example with HI 7030/HI 8030 turn the trimmer to read 12.88 mS/cm @ 25°C.
Page 17
• Turn the instrument on by pressing ON/OFF and select the appropriate range (e.g. 1999 mg/L). 1999mg/l ON/OFF • Tap the probe repeatedly on the bottom of the beaker and stir it to ensure that no air bubbles are trapped inside the sleeve. •...
Page 18
PROCEDURE FOR HI 9039 • Fill a beaker with 8 cm (3¼") of HI 7039 or HI 8039 conductivity calibration solu- tion at 5000 µS/cm @25°C (if possible fill two beakers and use one as rinse and the other for calibration). •...
Page 19
PROCEDURE FOR HI 933100 • Make sure that the conductivity probe is connected to the meter securely by align- ing the pins with the socket, pushing the plug in and tightening the threaded ring. • Dry the probe and leave it in the air. •...
Page 20
• If the reading is within 15% of the theoretical value, the meter automatically recognizes the buffer and the "BUF" symbol will stop flashing. • " " will disappear from the display µS when the reading is stable and "CON" °C symbol will start blinking.
Page 21
• Press CAL to display "84.0µS" together with the "CAL" symbol and an intermittent "BUF" indicator. µS °C • If the reading is within 15% of the theoretical value, then the meter automatically recognizes the value. • " " will disappear from the display µS when the reading is stable and "CON"...
C O N D U C T I V I T Y V E R S U S C O N D U C T I V I T Y V E R S U S C O N D U C T I V I T Y V E R S U S C O N D U C T I V I T Y V E R S U S C O N D U C T I V I T Y V E R S U S TEMPERATURE CHART...
FERTILIZER CALIBRATION FERTILIZER CALIBRATION FERTILIZER CALIBRATION FERTILIZER CALIBRATION FERTILIZER CALIBRATION ( ( ( ( ( HI HI HI HI HI 9039 9039 9039 9039 only 9039 only only only only ) ) ) ) ) • Turn the HI 9039 on by pressing ON/OFF. ON/OFF •...
TEMPERATURE COEFFICIENT TEMPERATURE COEFFICIENT TEMPERATURE COEFFICIENT TEMPERATURE COEFFICIENT TEMPERATURE COEFFICIENT (HI 933100 (HI 933100 HI 9635 HI 9635 only only only) ) ) ) ) (HI 933100 (HI 933100 (HI 933100 and and HI 9635 HI 9635 HI 9635 only only SETTING THE TEMPERATURE COEFFICIENT •...
• Measure the sample at a different temperature "t°C" which should be between 5°C to 10°C different from 25°C and note the conductivity reading as "Ct". • The temperature coefficient "ß" of the solution can be calculated with the following formula: (Ct - C25) ß...
TEMPERATURE CALIBRATION TEMPERATURE CALIBRATION TEMPERATURE CALIBRATION TEMPERATURE CALIBRATION TEMPERATURE CALIBRATION (HI 9039, HI 933100 (HI 9039, HI 933100 (HI 9039, HI 933100 and (HI 9039, HI 933100 (HI 9039, HI 933100 and HI 9635 HI 9635 HI 9635 HI 9635 only HI 9635 only only...
Page 28
• Immerse the conductivity probe into the 50°C (122°F) bath and wait for a couple of minutes for thermal equilibrium. 50 °C (122 °F) • Press CAL to display the "50.0°C", "CAL" symbol and an intermit- tent "BUF" indicator. " "...
L C D S Y M B O L S G U I D E µS Measurement in µS/cm (except HI 9034) Measurement in mS/cm (except HI 9034) For HI 9033, HI 9034 and HI 9130 only Out of range Low battery voltage For HI 9039, HI 933100 and HI 9635 only...
PROBE MAINTENANCE PROBE MAINTENANCE PROBE MAINTENANCE PROBE MAINTENANCE PROBE MAINTENANCE Rinse the probe with tap wa- ter after every series of measurements. If a more thor- ough cleaning is required, remove the PVC sleeve and clean the probe with a cloth or a nonabrasive detergent.
"BEPS" technology. BEPS, or Battery Error Preventive System, will detect a low power condition in the battery and display "V" (HI 9033, HI 9034 and HI 9130) or "LO BAT" (HI 9039, HI 933100 and HI 9635) on the LCD.
HI 8039L 5000 µS/cm (µmho/cm), 460 mL OTHER ACCESSORIES HI 721310 9V battery for HI 9033, HI 9034, HI 9039 and HI 9130 (10 pcs) HI 721308 1.5V AA size alkaline battery for HI 933100 and HI 9635 (10 pcs) CHECKTEMPC Electronic thermometer (range -50.0 to 150°C)
HI 731326 Small screwdrivers, length 90 mm, for calibration purposes (20 pcs) (for HI 9033, HI 9034 and HI 9130 only) HI 76301W 4-ring probe without temperature sensor, for Manual Temperature Compensation measurements with HI 933100 and HI 9635 HI 76302W 4-ring probe with temperature sensor, for Automatic...
CE DECLARATION OF CONFORMITY CE DECLARATION OF CONFORMITY CE DECLARATION OF CONFORMITY CE DECLARATION OF CONFORMITY CE DECLARATION OF CONFORMITY Recommendations for Users Before using these products, make sure that they are entirely suitable for the environ- ment in which they are used. Operation of these instruments in residential areas could cause unacceptable interfer- ences to radio and TV equipment.
Page 35
OTHER PRODUCTS FROM HANNA OTHER PRODUCTS FROM HANNA OTHER PRODUCTS FROM HANNA OTHER PRODUCTS FROM HANNA OTHER PRODUCTS FROM HANNA • CALIBRATION AND MAINTENANCE SOLUTIONS • CHEMICAL TEST KITS • CHLORINE METERS • DISSOLVED OXYGEN METERS • HYGROMETERS • ION SPECIFIC METERS (Colorimeters) •...
Page 36
HANNA LITERATURE HANNA LITERATURE HANNA LITERATURE HANNA LITERATURE HANNA LITERATURE Hanna publishes a wide range of catalogs and hand- books for an equally wide range of applications. The reference literature currently covers areas such as: • Water Treatment • Process •...
Need help?
Do you have a question about the HI 9033 and is the answer not in the manual?
Questions and answers