Hanna Instruments HI 731341 Instruction Manual

Hanna Instruments HI 731341 Instruction Manual

Fixed-volume pipette

Advertisement

Quick Links

HINTS FOR ACCURATE MEASUREMENTS
1. Use the pipette in a vertical position whenever possible.
2. Operate slowly and smoothly. Aspiration and dispensing should
been carried-out with the same speed.
3. The tip should be immersed in the sample liquid the minimum
necessary (2-4 mm). Greater depths may cause inaccurate
measures.
4. Change the tip whenever a different kind of sample is to be
aspired or when droplets remain on the inside wall of the tip
from the previous pipetting operation.
5. Work always with samples at room temperature.
PIPETTE MAINTENANCE
1. Liquid should never enter the pipette shaft. To avoid this:
• never use pipette without a tip
• do not invert pipette upside down or lay it flat when there
is a liquid in the tip
• press and release the push button slowly and smoothly.
2. When pipetting concentrated acids or alkalis or corrosive
solutions, pay particular attention to avoid contact of the
liquid with the pipette shaft, or it could be seriously damaged.
Extensive contact with corrosive fumes may result in premature
ageing of internal components of the pipette.
3. After use, store the pipette in its case or in a clean safe
place.
ACCESSORIES
HI 731350
tips for 200 L automatic pipette (25 pcs)
HI 731351
tips for 1000 L automatic pipette (25 pcs)
HI 731352
tips for 2000 L automatic pipette (4 pcs)
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.
Hanna Instruments Srl
V.le delle Industrie 12/A
35010 Villafranca (Padova)
http://www.hannainst.com
ITALY
Instruction Guide
Fixed-Volume Pipette
HI 731340
HI 731341
HI 731342
http://www.hannainst.com
Dear Customer,
Thank you for choosing a Hanna Instruments product.
This instruction guide will provide you with the necessary
information for the correct use of the pipette. Please read it
carefully before using the pipette. If you need additional technical
information, do not hesitate to e-mail us at tech@hannainst.com.
PRELIMINARY EXAMINATION
Please examine this product carefully. Make sure that no
damage has occurred during shipment. If there is any
damage, notify your Dealer.
Each pipette is supplied complete with:
• Two tips
• Instruction Guide
• Rigid carrying case
Note: Keep all packing material until you did verify that the
pipette works correctly. Any defective item must be
returned in its original packing.
GENERAL DESCRIPTION
HI 731340/1/2 is an air-displacement volumetric pipette
designed to measure and transfer fixed volumes of liquids
that needs to be precisely dosed.
The liquid is aspirated into deposable tips. The use of
disposable tips eliminate cross contamination of the liquid
samples.
To discard used tips, the pipette is equipped with a built-in
tip ejector to protect the user from contaminated tips.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instruments
Inc., Woonsocket, Rhode Island, 02895 , USA.

Advertisement

loading

Summary of Contents for Hanna Instruments HI 731341

  • Page 1 HI 731340/1/2 is an air-displacement volumetric pipette designed to measure and transfer fixed volumes of liquids Hanna Instruments reserves the right to modify the design, construction that needs to be precisely dosed. and appearance of its products without advance notice.
  • Page 2 Description: OPERATION Pushbutton 1. Mounting a tip: Tip ejector button first stop second stop Place a new tip on the pipette by pressing and gently rotating Pipette volume: the shaft into the tip. The tip will fit firmly the shaft. 200, 1000 or 2000 L Note: Never use the pipette without a tip.